25
PADnet
®
AND VEINnet
™
USER GUIDE
L-128 Rev B Jan 2013
BioMedix
®
•
Saint Paul, MN 55101
•
www.BioMedix.com
•
Toll Free: 888-889-8997
•
Fax: 651-762-4014
Arterial Testing
44. At the “0” minute interval immediately following
exercise, the software will prompt the user to take the
pressure of the arm with the highest recorded at-rest
pressure. Properly position the patient for testing. Click
“Start” or press 1 (Figure 1.33).
45. After the values display, click “Save”, or press 2
(Figure 1.33).
FIGURE 1.33 - Right Arm - “Stop”, “Save” and “Next” Buttons
46. Click “Next”, or press 4 (Figure 1.33).
47. Quickly move the hose to the Right Ankle Cuff and
click “Start”, or press 1 (Figure 1.34).
48. After the values display, click “Save”, or press 2
(Figure 1.34).
49. Click “Next”, or press 4 (Figure 1.34).
FIGURE 1.34 - Right Ankle - “Start”, “Save” and “Next” Buttons
50. Quickly move the hose to the Left Ankle cuff and
repeat the pressure taking process using steps 44–49.
51. After ankle pressures are complete, record the PVR
of both ankles and the arm with the highest pre-exercise
pressure.
52. Click “Start” to record the Right Ankle PVR (Figure
1.35).
FIGURE 1.35 - Right Ankle - “Start” Button
53. After a full screen of consistent quality waveforms
has appeared, click “Stop” (“Start” button changes to
“Stop”), or press 1 (Figure 1.36, next page).
Post-exercise at: 0 minutes