37. After at-rest ABI values and PVR waveforms are
recorded, click “Post-Excercise Test...” from the Test
Summary screen (Figure 1.29).
FIGURE 1.29 - “Post Exercise Test...” Button
Note: After Post-Exercise test results have been saved,
the initial at-rest test will no longer be accessible for
changes.
38. Remove Above Knee and Calf Cuffs, and have the
patient walk on a motorized treadmill until the onset of
symptoms, or for up to five minutes.
39. To start the exercise timer, click “Start”, or press 1
(Figure 1.30).
FIGURE 1.30 - Exercise Period - “Start” Button and Notes Field
40. Under “Notes”, record the treadmill speed (2 mph),
grade (12%) and any reported symptoms (Figure 1.31).
41. At the onset of symptoms, click “Stop” (“Start”
button changes to “Stop”) (Figure 1.31).
FIGURE 1.31 - Exercise Period - “Stop” Button
42. Click “Save” (Figure 1.32).
43. Click “Next”, or press 4 to move to the test screen
(Figure 1.32).
FIGURE 1.32 - Exercise Period - “Save” and “Next” Buttons