Recommended separation distances between portable and mobile RF communication devices and the PRD-7130100C,
PRD-7130110C, PRD-7130120C or PRD-7130130C
The PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-7130130C models are designed to be used in electromagnetic
environments in which radiated RF emissions are controlled. PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-
7130130C customers and users can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communication devices (emitters) and the PRD-7130100C, PRD-7130110C, PRD-7130120C or PRD-
7130130C, as recommended below and in accordance with the maximum emitted power of the communication devices in
question.
Maximum assigned output power
of the emitter
W
Separation distance according to emitter frequency
150 kHz to 80 MHz
d = 1.17 √ P
80 MHz to 800 MHz
d = 1.17 √ P
800 MHz to 2.5 GHz
d = 2.34 √ P
0.01
0.1
0.37m
0.37m
0.74m
1
10
100
In the case of an emitter whose assigned maximum output power is not given above, the recommended separation
distance d in metres (m) can be estimated using the applicable equation for the emitter frequency, where P is the maximum
characteristic power output of the emitter in Watts (W) according to its manufacturer.
NOTE 1 At 80 MHz and at 800 MHz, the separation distance for the highest range of frequencies is applicable.
NOTE 2 These directives might not be applicable to all situations. Electromagnetic propagation is affected by absorption and
reflections of structures, objects and persons.
Directives and manufacturer declaration - electromagnetic immunity
The PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-7130130C models are designed to be used in the
electromagnetic environment specified below. PRD-7130100C, PRD-7130110C, PRD-7130120C and PRD-7130130C
customers or users must ensure that their devices are used in such an environment.
Immunity test
Test level
CEI 60601
Level of
compliance
Electromagnetic environment - directives
Electrostatic
discharges
(ESD)
CEI 61000-4-2
± 6 kV on contact
± 8 kV into the air
± 6 kV on contact
± 8 kV into the air
The floors must be made of wood, concrete
or ceramic tiles. If the floors are covered with
synthetic material, relative humidity must be
at least 30%.
Fast
transient bursts
CEI 61000-4-4
± 2 kV for electrical
supply lines
± 1 kV for
input/output lines
± 2 kV for electrical
supply lines
± 1 kV for input/output
lines
The quality of the electrical power supply
network must equate to that in a typical
commercial or hospital environment.
Transient
overvoltages
CEI 61000-4-5
± 1 kV between
phases
± 2 kV between
phase and earth
± 1 kV between phases
± 2 kV between phase
and earth
The quality of the electrical power supply
network must equate to that in a typical
commercial or hospital environment.
Voltage dips, short
outages and voltage
fluctuations in the
electrical power supply
input lines
CEI 61000-4-11
<5% UV
(>95% UV dip) for
0.5 cycles
<40 % UV
(60 % UV dip) for 5
cycles
<70% UV
(30% UV dip) for 25
cycles
<5% UV
(>95% UV dip) for 5
seconds
<5% UV
(>95% UV dip) for 0.5
cycles
<40 % UV
(60 % UV dip) for 5
cycles
<70% UV
(30% UV dip) for 25
cycles
<5% UV
(>95% UV dip) for 5
seconds
The quality of the electrical power supply
network must equate to that in a typical
commercial or hospital environment. If a
PRD-7130100C, PRD-7130110C, PRD-
7130120C or PRD-7130130C user requires
continuous operation during electrical power
supply network outages, we recommend
connecting the PRD-7130100C, PRD-
7130110C, PRD-7130120C or PRD-7130130C
to an uninterruptible power supply or
battery.
Magnetic field at the
electrical network
frequency (50/60 Hz)
CEI 61000-4-8
3 A/m
3 A/m
Magnetic fields at the electrical network
frequency must be at levels that are
characteristic of a typical commercial or
hospital environment.