Declaration of Conformity
Emissions
manufacturer’s declaration electromagnetic emissions
The Balance System SD is intended for use in the electromagnetic environment specified below. The
customer or the user of the Balance System SD should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment
RF emissions CISPR 11
Group 1
The Balance System SD generates RF energy
only for its internal functions. Therefore, its RF
emission is very low and is not likely to cause
any interference in nearby electronic equipment
RF emissions CISPR 11
Class A
The Balance System SD is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network supplying buildings
used for domestic purposes.
Harmonic distortion
Class A
EN 61000-3-2
Voltage fluctuations
Complies
and flickerEN 61000-3-3
Immunity
manufacturer’s declaration electromagnetic immunity
The Balance System SD is intended for use in the electromagnetic environment specified below. The
customer or the user of the Balance System SD should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2
IEC 60601-1-2
Electromagnetic
Test level
Compliance level
environment – guidance
Electrostatic discharge
± 6 kV contact
Contact ± 6 kV
Floor should be wood,
concrete or ceramic tiles.
If floor is covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical fast
± 2 kV for power lines
Power ± 2 kV
Mains power quality
transients/burst
± 1 kV for
Signal ± 1 kV
should be that of a
IEC 61000-4-4
input/output lines
typical commercial or
hospital environment
Surge
± 1 kV differential mode
± 1 kV diff. mode
Mains power quality
IEC 61000-4-5
± 2 kV common mode
± 2 kV com. mode
should be that of a
typical commercial or
hospital environment
CONTENTS
APPENDIX E
— E-2 —
Содержание 950-440
Страница 47: ...Figure 8 11 A sample Athlete Single Leg Stability Results report CONTENTS 8 11 TESTING ...
Страница 64: ...9 3 Limits of Stability Single Leg Test Report unilateral stance 14 GENERAL MAINTENANCE REPORTS 9 6 ...
Страница 65: ...9 4 Limits of Stability Test Report bilateral stance 15 SERVICE PROCEDURES 9 7 REPORTS ...
Страница 66: ...9 5 Athlete Single Leg Stability Test Report CONTENTS REPORTS 9 8 ...
Страница 105: ... R 1 REPLACEMENT REPLACEMENT ...
Страница 106: ...REPLACEMENT REPLACEMENT R 2 ...
Страница 107: ...REPLACEMENT PARTS R 3 REPLACEMENT ...
Страница 108: ...REPLACEMENT PARTS REPLACEMENT R 4 ...
Страница 109: ...REPLACEMENT PARTS R 5 REPLACEMENT ...
Страница 110: ...REPLACEMENT PARTS REPLACEMENT R 6 ...
Страница 111: ...REPLACEMENT PARTS R 7 REPLACEMENT ...
Страница 112: ...REPLACEMENT PARTS REPLACEMENT R 8 ...