Biochrom Anthos Fluido-2 Microplate Washer v2.0
Page 74
Close fuse-carrier.
Turn device on and check function. In case of malfunction, call a
service technician.
11.4
Cleaning the instrument
This instrument is a precision instrument and the requires regular cleaning
to ensure the continued precision.
Liquid spillage
If any liquid is spilled on the instrument, it should be IMMEDIATELY
removed so that the liquid does not attack the surface of the housing.
Regular cleaning
The housing of the instrument should be cleaned regularly with a mild
household cleaning agent.
Warning: Do not use aggressive solutions!
The area surrounding the plate carrier should be cleaned with a dry cloth.
Fittings and tubing
The usual wash buffer tends to crystallize easily on air and can block
tubing and fittings then. Use the automatic rinse program daily and do not
leave the tubing filled with wash buffer for an extended period of time.
11.5
Disinfection Procedure
If the laboratory has no specific disinfection procedure, the following
procedure should be used to disinfect the instrument.
The instrument should be disinfected using a suitable disinfection
solution.
1. Disconnect the instrument from the mains power supply.
2. Carefully wipe all the outside surfaces of the instrument and the plate
area with a wad of cotton wool that has been soaked in the disinfection
solution.
Note: Ensure that disposable gloves are worn.