Biochrom Anthos Fluido-2 Microplate Washer v2.0
Page 10
Operating voltage:
90-130V, 180-250VAC (autosensing),
frequency range:
47 - 63 Hz
4.2.3
Interfacing
Serial interface (RS232 9-pin male connector) for remote control of basic
wash functions, instrument status and setup by external PC (with null-
modem serial cable used for PC connection).
4.2.4
Options
liquid level sensors for wash and waste bottles
8- and 12-way manifold for 96-well plates
16-way manifold for 384-
Serial interface cable
4.3
Cleaning
This instrument is a precision instrument and it requires regular cleaning
to ensure the continued precision.
Liquid spillage
If any liquid is spilled on the instrument, it should be IMMEDIATELY
removed so that the liquid does not attack the surface of the housing.
Regular cleaning
The housing of the instrument should be cleaned regularly with a mild
household cleaning agent.
Warning: Do not use aggressive solutions!
The area surrounding the plate carrier should be cleaned with a dry cloth.
Fittings and tubing
The usual wash buffer tends to crystallize easily on air and can then block
tubing and fittings. Use the automatic rinse program daily and do not leave
the tubing filled with wash buffer for an extended period of time.
4.4
General features
The Fluido 2 incorporates state of the art technology and software both
necessary to allow optimum performance and easy use.
The Fluido 2 is designed around your application.