![Biocare EverHeart EH-01 Скачать руководство пользователя страница 5](http://html.mh-extra.com/html/biocare/everheart-eh-01/everheart-eh-01_user-manual_2753374005.webp)
2
Product specification
3
Model EH-01
Product Specification
155mm(H)*140mm(D)*110mm(W) (w/o chamber)
Weight 1.3kg (w/o chamber fitted)
1.6kg (w/ chamber fitted and filled with water)
Electrical Rating Frequency
:
47~63Hz
Voltage
:
AC100 ~240V~/1.8A max./ 60W
Fuse ”T24M105ASR2”
Overheating protection The machine cuts off its power when temperature over 105
℃
Heater Control Adjustable 1~9 level
Setting Range 49
℃
~73
℃
Max temperature 73
℃
Disconnection device Power On/Off switch from the rear side
Max Operating Pressure 20kpa
Heat equivalent time Less then 20 minutes
●
Transportation and Storage environmental condition
Transport Condition -10
℃
to +55
℃
( RH5%~95%)
Storage Condition -20
℃
~ +65
℃
( RH5%~95%)
Standard Compliance EN/IEC-60601-1 , ISO8185
Heater 24V/60 W ; -60
℃
~150
℃
●
Operating environment
Temperature 5
℃
to 40
℃
( 41
℉
~104
℉
)
Humidity <80%(non-condensing)
Atmospheric Pressure 619 to 775mmHg(825 to 1033 hPa)
Electrical/Magnetic contact Non Effect