background image

2

1.2    Definitions  -  Symbol

Symbol

Symbol

Explanation

Explanation

Type B                                                                                        Caution

Manufacture Date                                                                 Protective Earth

              

               

  

Alternating current

                        

Caution : Electrostatic sensitive device

                                                                                                                     

Caution :   
Hot surface may exceed 74

                                                     

Caution: Electrical Shock Hazard. 
Refer to qualified service personnel.

Consult instructions for use

Содержание EverHeart EH-01

Страница 1: ...Respiratory Humidifier User Manual Model EH 01 ...

Страница 2: ...1 2 Definitions 2 Product specifications 3 General Information 4 Explanation of controls and indicators 5 Assembling 6 Operation 6 1 Basic Step 7 Daily Checking 8 Product Performance 9 Routine Maintenance 10 Service Procedure 1 2 3 4 5 6 7 8 8 9 10 ...

Страница 3: ...as flow remains at a constant level 1 2 Intended Use Disposable heated humidification chamber is intended for use to hold water required to humidify the air being delivered to patients This device is not intended for patients whose upper airways have been bypassed Caution This humidifier did not have temperature monitor function in patient side Please add an external temperature probe when needed ...

Страница 4: ... Explanation Type B Caution Manufacture Date Protective Earth Alternating current Caution Electrostatic sensitive device Caution Hot surface may exceed 74 Caution Electrical Shock Hazard Refer to qualified service personnel Consult instructions for use ...

Страница 5: ...ontrol Adjustable 1 9 level Setting Range 49 73 Max temperature 73 Disconnection device Power On Off switch from the rear side Max Operating Pressure 20kpa Heat equivalent time Less then 20 minutes Transportation and Storage environmental condition Transport Condition 10 to 55 RH5 95 Storage Condition 20 65 RH5 95 Standard Compliance EN IEC 60601 1 ISO8185 Heater 24V 60 W 60 150 Operating environm...

Страница 6: ...e personnel authorized to do so by the manufacture Use only parts supplied or approval by BioCARE 3 Compliance with the local electrical installation regulations 4 Maintenance of grounding integrity by connection to a hospital grade receptacle Always disconnect supply before servicing Caution 1 This product is intended for use by a qualified medical practitioner and such people should ensure that ...

Страница 7: ...level Temperature rising key Press one time to higher setting temperature level Temperature decreasing key Press one time to lower setting temperature level Rear side On Off Switch Power On o Power Off Hanging slot This slot can help use to fasten the machine on the rod or ventilator On Off switch Hanging slot Display widow Unlock Key Temperature rising key Temperature decrease key Heating Alumina...

Страница 8: ...ngth of 24 inches between the reservoir and the patient connection 5 Connect the patient end air way to the chamber outlet side 6 Strongly recommend add a small temperature in the patient side to monitor the temperature 7 Connect the air way between patient and ventilator segment 8 Now the humidifier is ready to use Caution Do not fill the chamber above the maximum level Liquid could enter the bre...

Страница 9: ...on Do not touch the heater plate the temperature on the surface may exceed 73 Always make sure that gas flows through the humidifier before connecting circuit to patient A possible explosion hazard may exists if use in the presence of flammable anesthetics Switch the humidifier off if gas flow is stopped or interrupted Use BioCARE or other compatible chamber Ensure that the water level in the cham...

Страница 10: ...ambient temperature gas flow from ventilator and breathing circuit setting Level setting 1 2 3 4 5 6 7 8 9 49 52 55 58 61 64 67 70 73 32 33 34 35 36 37 38 38 39 27 27 28 29 30 30 31 31 32 Heating plate temperature Chamber outlet wye connector patient site Caution This device may not be able to function appropriately or provide sufficient humidity if operated outside the intended specified ambient ...

Страница 11: ...leaned by using dry damp cloth 9 2 Cancel Check the humidifier case for cracking replace if necessary 9 3 Cleaning instruction Please using the following washing up solution on the everHeart humidifier to avoid the damage on the plastic or metal parts Method 1 Disconnect the humidifier from any electrical outlet 2 Clean the humidifier with one of the following recommended disinfectants using a dam...

Страница 12: ...e proper operation Where screws and bolts have been removed from the product do not use excessive force when re fastening 10 2 Warranty Repair Please contact your local distributor for warranty repair 10 3 Non warranty Repair Do not disassemble the everHeart humidifier by non qualified engineers or without authorization 10 4 Discard Please discard everHeart and its accessory carefully according to...

Страница 13: ... 2015 BioCARE Taiwan All Rights Reserved 9010125 REV 04 ...