28
System Utilities – Editors
Protocol Editor
�
Note: Only accessible under ‘Supervisor’ level security log-in.
�
The ‘Protocol Editor’ window displays all protocols currently programmed into the software. The protocols are separated
into groups by detection system. A user may toggle between the different protocol templates by selecting their respective
tabs found at the top of the ‘Protocol Editor’ window.
�
The ‘Protocol Editor’ allows the user logged-in under the “
Supervisor” setting only to alter 5 parameters within programmed
protocols.
�
Note: Some features in the Protocol Editor are reserved for future use.
DS2 (Dewax Solution 2) Option
To allow for the substitution of the DS2 component of the deparaffinization kit with Buffer, an Enzyme or a Block.
�
Some antibodies are sensitive to a component of DS2. In this instance, DS Buffer may be desired instead.
�
If an antibody requires enzyme digestion before antigen retrieval, then DS Enzyme would be desired.
�
If an antibody requires a block before antigen retrieval, then DS Block would be desired.
�
It is recommended to consult a Biocare representative if making changes to the DS2 setting.