G.4. pH Regulation Menu
G.4.1. Manual injection
Menu
Functions
Possible
settings
Default setting
Instructions
PH Regulation
Manual Injection
•
Priming
of
the
peristaltic pump and
filling
of
semi-rigid
pipes.
•
pH corrector injection.
•
Means of checking the
correct operation of the
peristaltic pump.
From
30
seconds to 10
minutes,
in
increments of 30
seconds.
1 min
•
To start injecting :
Confirm the duration setting.
(The peristaltic pump is running,
and a timer countdown is
displayed in real time.)
•
To take a break, and to restart
the injection :
Press on
.
•
To stop the injection :
Press on
.
G.4.2. Activation/deactivation of pH regulation
Menu
Possible settings
Default setting
PH Regulation
Mode XXX
•
ON
•
OFF
ON
G.4.3. Calibration of the pH measurement
Menu
Possible settings
Default setting
PH Regulation
Adjust
From 6,5 to 7.5, in increments of 0.1.
Measurement displayed
G.4.4. Calibrating the pH probe
Important advance information
The original pH probe is already calibrated. It is therefore not necessary to carry out calibration of the pH probe
when putting the equipment into service for the first time.
However, it is imperative to carry out a calibration of the pH probe at the beginning of each
season when returning to service, and after each probe replacement.
1)
Open the pH 7 and pH 10 calibration solutions (use only single-use calibration solutions).
2)
Turn off the filtration (and therefore the electronics unit).
3) If the probe is already installed :
a)
Remove the probe from the probe holder, without disconnecting it.
b)
Remove the probe holder nut and replace it with the stopper supplied.
If the probe is not already installed :
Connect the probe to the electronics unit.
4)
Turn on the electronics unit.
5)
Go to the «
pH Regulation - Calibration
» menu.