Chapter 3: Operation
Overview
IMPORTANT!
Do not turn on the instrument until the
carrier shipping block has been removed.
The ELx800 features a 25-pad keypad and a 2-line x 24-character LCD display, allowing
you to access the reader’s program menus and print test results. The reader’s bidirectional
serial port allows computer control of the instrument, and provides the means for
downloading additional assay definition files to the instrument. This chapter describes the
operation of the
open
(configurable)
assays onboard the ELx800
.
Recommendations for Achieving Optimum
Performance
Microplates should be perfectly clean and free of dust or bottom scratches. Use new
microplates from sealed packages. Do not allow dust to settle on the surface of the
solution; use microplate covers when not reading the plate. Filter solutions to remove
particulates that could cause erroneous readings.
Although the readers support standard flat, U-bottom, and V-bottom microplates,
optimum performance is achieved with optically clear, flat-bottomed wells.
Non-uniformity in the optical density of the well bottoms can cause loss of accuracy,
especially with U- and V-bottom polyvinyl microplates. Check for this by reading an
empty microplate. Dual-wavelength readings can eliminate this problem, or bring the
variation in density readings to within acceptable limits for most measurements.
Inaccuracy in pipetting has a large effect on measurements, especially if smaller volumes
of liquid are used. For best results, use at least 100
µ
l per well in a 96-well plate and 25
µ
l
in a 384-well plate.
Dispensing solution into 384-well plates often traps air bubbles in the wells, which may
result in inaccurate readings. A dual-wavelength reading method usually eliminates these
inaccuracies; however, for best results, remove the air bubbles by degassing the plate in a
vacuum chamber before reading.
The inclination of the meniscus can reduce reading accuracy in some solutions, especially
with small volumes. Agitate the microplate before reading to help bring this problem
within acceptable limits. Use Tween® 20, if possible (or some other wetting agent) to
normalize the meniscus. Some solutions develop menisci over a period of several minutes.
This effect varies with the brand of microplate and the solution composition. As the center
of the meniscus drops and shortens the light path, the density readings change.
Bio-Tek Instruments, Inc.
Содержание ELx800
Страница 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Страница 2: ......
Страница 4: ......
Страница 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Страница 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Страница 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Страница 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Страница 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Страница 192: ...164 Appendix C Report Format Figure 31 Samples with calls on Matrix Report Bio Tek Instruments Inc...
Страница 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Страница 194: ...166 Appendix C Report Format Figure 33 Samples with calls on Column Report Bio Tek Instruments Inc...
Страница 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Страница 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Страница 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Страница 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Страница 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...