AMPLIFICATION

CFX96 Touch

 Real-Time PCR System 

Installation Quick Guide

Unpacking the Instruments

The CFX96

 optical reaction module and the C1000 Touch

 

thermal cycler chassis are shipped in separate containers. 
Unpack each box carefully. The included power cable, USB 
cable, and CFX Manager

 software CD are required for 

installation. The CFX96 optical reaction module may also be 
placed on a previously purchased C1000 Touch thermal cycler 
for a quick upgrade to a real-time PCR detection system.

Important:

 Carefully read the safety information and 

instrument operating specifications provided in the CFX96 
Touch

 and CFX384 Touch

 Real-Time PCR Detection 

Systems Instruction Manual before using the system.

Setting Up the CFX96 Touch Real-Time PCR System

The CFX96 Touch real-time PCR detection system should 
be installed on a clean, dry, level surface with sufficient 
cool airflow to run properly.

To insert the CFX96 optical reaction module into the 
reaction module bay of the C1000 Touch chassis:
1.   Place the C1000 Touch chassis in a suitable location 

with the locking bar down.

2.   Remove the instructions overlay from the C1000 Touch 

chassis. Lift the CFX96 module using the handle 
indents above the side air vents, as shown in Figure 1. 

3.   Position the CFX96 module in the reaction module bay 

of the C1000 Touch chassis, leaving about 2 cm of 
space in the front.

4.   Reach around the thermal cycler and pull up the 

locking bar (Figure 2) until it is flush with the sides of 
the module bay. This action moves the reaction module 
forward, locking it into place (Figure 3).

5.   Use the supplied power cord to plug the chassis into  

an appropriate electrical outlet. 

6.   Turn on the CFX96 Touch system using the power 

switch on the back of the chassis.

Removing the Red Shipping Screw

The CFX96 optical reaction module is shipped with a  
red screw in place in the inner lid in the hole corresponding 
to well A1 (see Figure 4). Follow the instructions on the 
C1000 Touch screen to remove this shipping screw before 
beginning operation.

Fig. 3. The CFX96 module positioned on 
the C1000 Touch chassis.

Locking bar

Fig. 1. Lifting the CFX96 optical reaction module  
onto the C1000 Touch thermal cycler chassis.

Fig. 2. Locking the CFX96 module into place. 

C1000 Touch thermal  
cycler chassis

CFX96 optical  
reaction module

Fig. 4. Red shipping screw included with  
the CFX96 optical reaction module.