BIO RAD CFX96 Touch Скачать руководство пользователя страница 4

Life Science

Group

10-1805    0311    Sig 0211

10021418 Rev A     US/EG

Bio-Rad 

Laboratories, Inc.

Web site 

www.bio-rad.com  

USA 

800 424 6723  

Australia 

61 2 9914 2800  

Austria 

01 877 89 01  

Belgium 

09 385 55 11 

Brazil 

55 31 3689 6600  

Canada 

905 364 3435  

China 

86 21 6169 8500  

Czech Republic 

420 241 430 532  

Denmark 

44 52 10 00  

Finland 

09 804 22 00  

France 

01 47 95 69 65  

Germany 

089 31 884 0  

Greece 

30 210 777 4396  

Hong Kong 

852 2789 3300  

Hungary 

36 1 459 6100  

India 

91 124 4029300  

Israel 

03 963 6050  

Italy 

39 02 216091  

Japan 

03 6361 7000  

Korea 

82 2 3473 4460  

Malaysia 

60 3 2117 5260  

Mexico 

52 555 488 7670  

The Netherlands 

0318 540666  

New Zealand 

64 9 415 2280  

Norway 

23 38 41 30  

Poland 

48 22 331 99 99  

Portugal 

351 21 472 7700  

Russia 

7 495 721 14 04  

Singapore 

65 6415 3170  

South Africa

 27 861 246 723  

Spain 

34 91 590 5200  

Sweden 

08 555 12700  

Switzerland 

061 717 95 55  

Taiwan 

886 2 2578 7189 

 Thailand 

66 2 6518311   

United Kingdom 

020 8328 2000

Analyzing a Data File

A stand-alone run data file (.zpcr) needs to be converted into a data analysis file  
(.pcrd) by CFX Manager software in order to be analyzed. Create a data analysis file  
from a stand-alone run using one of the following options:

n

   Click and drag the .zpcr file over the main software window

n

   Select 

File > Open > Stand-alone Run

 from the main software window menu options

 After the file has been created, click the 

Select Plate

 button to import a plate file.  

The software will use this plate information to name and analyze the data file. The scan 
mode and plate size for the plate file must match the run settings used for the run. 
Refer to the CFX96 Touch and CFX384 Touch systems instruction manual for detailed 
information about data analysis.

Setting Up Email on the C1000 Touch Chassis

To configure the outgoing email from the C1000 Touch chassis, follow these instructions:
1.   If you have not already done so, in the home screen, select 

Tools

, then touch the  

Admin

 button to log in as the administrator. If no password has been set, the Admin 

tools list appears without a log-in prompt.

2.  In the Admin tools list, select 

Email Settings

 (Figure 13).

3.   Select a mail server from the drop-down menu in the Outgoing Mail (SMTP) Server 

Settings window or create a new one by touching the 

New Server

 button (Figure 14). 

Make this server the default by touching the 

Set As Default

 box.

4.   To create a new mail server, enter the mail server address and mail server port in  

the appropriate fields. A network administrator may need to be contacted to obtain 
this information.

5.  Enter the user name and password for the selected mail server.
6.   To verify that the Outgoing Mail (SMTP) Server Settings are correct, a test email can 

be sent by touching the 

Test Email

 button. 

7.  Enter the email address and an attachment size, then touch 

Send Email

 (Figure 15).

Windows and Windows Vista are trademarks of Microsoft Corporation.
Notice regarding Bio-Rad thermal cyclers and real-time systems: 
Purchase of this instrument conveys a limited non-transferable immunity from suit for the purchaser’s own internal research and development and for use in human 
in vitro diagnostics and all other applied fields under one or more of U.S. Patent Numbers 5,656,493; 5,333,675; 5,475,610 (Claims 1, 44, 158, 160–163, and 167 
only); and 6,703,236 (Claims 1–7 only), or corresponding claims in their non-U.S. counterparts, owned by Applera Corporation. No right is conveyed expressly, by 
implication or by estoppel under any other patent claim, such as claims to apparatus, reagents, kits, or methods such as 5' nuclease methods. Further information 
on purchasing licenses may be obtained by contacting the Director of Licensing, Applied Biosystems, 850 Lincoln Centre Drive, Foster City, California 94404, USA.
Bio-Rad’s real-time thermal cyclers are licensed real-time thermal cyclers under Applera’s United States Patent Number 6,814,934 B1 for use in research, human in 
vitro diagnostics, and all other fields except veterinary diagnostics.
Bio-Rad’s real-time thermal cyclers are covered by one or more of the following U.S. patents or their foreign counterparts owned by Eppendorf AG: U.S. Patent 
Numbers 6,767,512 and 7,074,367.

Fig. 14. Outgoing Mail (SMTP) Server 
Settings window.

Fig. 15. Test Email window.

Fig. 13. Tools window following Admin 
log-in.