Defining New Programs
50 Bio-Plex Pro and Bio-Plex Pro II Wash Stations
Aspirating Modes
To improve wash efficiency and reduce the residual volume, the manifold must be
positioned so that the aspirating needles are correctly positioned in the wells for
round-bottom, V-shaped bottom, or flat-bottom well microplates.
Normal AspirationMode
For round-bottom or V-shaped bottom wells, the aspirating needles are placed in the
middle of the wells. Only one aspiration position can be selected.
Round Bottom Wells
'V' Shaped Wells
Normal Aspirating Positions
Crosswise Aspiration Mode
With flat bottom wells, the instrument can perform crosswise aspiration using two
aspiration positions per well.
The aspirating needles are set at two positions on the bottom of the wells
(front edge and back edge).
Aspirate Y Offset 1
Aspirate Y Offset 2
Front edge
Back edge
Crosswise Aspiration for Flat Bottom Wells
Содержание Bio-Plex Pro
Страница 2: ...ii Bio Plex Pro and Bio Plex Pro II Wash Stations...
Страница 4: ...iv Bio Plex Pro and Bio Plex Pro II Wash Stations...
Страница 10: ...Basic Operation and Maintenance 10 Bio Plex Pro and Bio Plex Pro II Wash Stations...
Страница 37: ...About the Instrument Bio Plex Pro and Bio Plex Pro II Wash Stations 37 Bio Plex Pro Dimensions...
Страница 38: ...About the Instrument 38 Bio Plex Pro and Bio Plex Pro II Wash Stations Bio Plex Pro II Dimensions...
Страница 44: ......
Страница 46: ...Defining New Programs 46 Bio Plex Pro and Bio Plex Pro II Wash Stations...
Страница 47: ...Defining New Programs Bio Plex Pro and Bio Plex Pro II Wash Stations 47...
Страница 64: ......
Страница 79: ...Bio Plex Pro and Bio Plex Pro II Wash Stations 79 Declaration of Conformity...
Страница 81: ...Bio Plex Pro and Bio Plex Pro II Wash Stations 81...
Страница 82: ...82 Bio Plex Pro and Bio Plex Pro II Wash Stations...