User Manual

 

 

             

 

 

 

Bio-Optica Milano S.p.A. Via San Faustino 58 - 20134 Milan  

Phone +39 02.21.27.13.1 - Fax Italy +39 02.21.53.000 - Fax 39 02.21.54.155 

www.bio-optica.it

 – 

[email protected]

 – 

[email protected]

 

 

 

5

 

Publication date: 02/03/2020 

Rev. 00 

1

 

IMPORTANT INFORMATION  

1.1  PRECAUTIONS FOR USE 

Before using the instrument, carefully read the instructions and warnings contained in this manual and keep 
them for further reference. They provide important information regarding the functions and safety in installing, 
using and maintaining the instrument. 
Bio-Optica Milano S.p.A. cannot be held liable for any damage resulting from improper or incorrect use and for 
failure to comply with the provisions of this manual and the current safety standards. 

 

1. After removing the  packaging, make sure that the instrument is intact without  visible damage that may  have been 
caused by transportation. 
 
2. Before connecting the instrument, make sure that the data on the nameplate corresponds to that of the power supply. 
 
3. This instrument must only be used for the purpose for which it was expressly designed, that is, as a paraffin dispenser 
for laboratory use. Any other use is to be considered improper and therefore hazardous. 
 
4. The instrument must only be used by authorized and professionally qualified personnel. 
 
5.  Periodic  annual  maintenance  must  be  carried  out  only  by  qualified  personnel  authorized  by  Bio-Optica.  For 
information, please contact the Technical Assistance Service at Tel. 02-21271310. 

 

6. The electrical safety of this instrument is ensured only when it is correctly connected to an efficient grounding system 
as required by current electrical safety standards. It is necessary to check this fundamental safety prerequisite, and if in 
doubt, please request a thorough check of the system. The instrument is equipped with a power supply cable having 2 
wires + earth to be connected to the power outlet. 
 
7. Do not remove the frame or its parts during operation. Turn off the instrument and disconnect it from the power outlet 
before opening it. This operation must only be carried out by authorized and professionally qualified personnel. 
 
8.  To  eliminate  the  risk  of  instrument  malfunction,  the  instrument  must  operate  in  an  environment  free  of  strong 
electromagnetic fields; this means that transmitters such as cell phones should not be used near the instrument. 
In case of serious malfunction, turn off the instrument and contact the Technical Assistance Service. 
 
9. All scraps and wastes, both infectious and radioactive, deriving from the work cycle of the instrument must be disposed 
of in compliance with the relevant laws in force. 
 

   

This  equipment  is  marked  with  the  symbol  shown  on  the  side,  in  accordance  with  the  European  Parliament 

Directive 2002/96/EC and subsequent amendments regarding WEEE (Waste Electrical and Electronic Equipment). This 
means that it is forbidden to dispose of this appliance as normal waste;  it must instead be taken to a special WEEE 
collection center authorized and set up by the Public Administration. 
 
10. The contents of this manual are subject to change without notice. 
 
11. Please find enclosed the declaration of conformity. 
 
 
 

Содержание BEC150

Страница 1: ...raffin embedding station In vitro diagnostic medical device CND code W0202059002 Bio Optica code 40 200 002 Reference number code 1831667 Always keep this manual near the instrument Read thoroughly be...

Страница 2: ...ica Milano S p A Via San Faustino 58 20134 Milan Phone 39 02 21 27 13 1 Fax Italy 39 02 21 53 000 Fax Export 39 02 21 54 155 www bio optica it info bio optica it claims bio optica it 2 Publication dat...

Страница 3: ...ructions or specific information provided in this manual Instructions drawings illustrations and any other technical or general information in this user manual are not binding with regard to our produ...

Страница 4: ...1 2 SYMBOLS AND THEIR MEANING 6 1 3 PROPER USE 7 1 4 USER GROUP 7 1 5 TYPE OF INSTRUMENT 7 1 6 SAFETY WARNINGS 8 1 7 DANGER WARNINGS 8 2 DESCRIPTION 9 2 1 MAIN COMPONENTS 9 2 2 MAIN CHARACTERISTICS 10...

Страница 5: ...02 21271310 6 The electrical safety of this instrument is ensured only when it is correctly connected to an efficient grounding system as required by current electrical safety standards It is necessa...

Страница 6: ...20 Rev 00 1 2 SYMBOLS AND THEIR MEANING Symbol for CATALOG NUMBER Symbol for SERIAL NUMBER Symbol for ALTERNATING CURRENT Symbol for FUSE Symbol for CONSULT THE INSTRUCTIONS FOR USE Symbol for CE MARK...

Страница 7: ...he required temperature Fill the molds containing the tissue samples with paraffin Heat and maintain the temperatures of the embedding cassettes with samples and molds Set and manage temperatures sepa...

Страница 8: ...er use and in a state of perfect technical safety Faults that can compromise safety must be eliminated immediately 1 7 DANGER WARNINGS The safety devices supplied on this instrument by the manufacture...

Страница 9: ...x Export 39 02 21 54 155 www bio optica it info bio optica it claims bio optica it 9 Publication date 02 03 2020 Rev 00 2 DESCRIPTION 2 1 MAIN COMPONENTS 1 Tank 2 Display 3 On off button 4 Flow adjust...

Страница 10: ...e consistently kept at a proper temperature so as to ensure the constant maintenance of the desired working temperature Dimensional characteristics Dimensions WxDxH 560 x 605 x 405 mm Paraffin tank ca...

Страница 11: ...display Detection of paraffin level via integrated level sensors Detection of presence of paraffin collection cassettes by means of presence sensors so as to increase workplace safety Heated and remo...

Страница 12: ...e instrument grip it only at the lower base of the housing at the front or rear and pull it out Installation Place the instrument on a flat and stable work bench and connect it to the power outlet 230...

Страница 13: ...a it claims bio optica it 13 Publication date 02 03 2020 Rev 00 4 USE 4 1 INSTRUMENT PARTS Paraffin tank 1 The paraffin tank has a capacity of 4 liters The paraffin temperature can be adjusted between...

Страница 14: ...h the blue knob 8 located above the nozzle Workspace 3 The temperature of the work area can be adjusted between 55 C and 75 C it includes the embedding area the forceps holder 5 and the cold spot 4 A...

Страница 15: ...facilitate work for ambidextrous operators or to work with forceps and pestle at the same time Trays for molds or cassettes 12 Depending on the preferred working direction the two heated trays can be...

Страница 16: ...n it will be in stand by mode The control panel provides an overview of the machine s operating status The upper part shows a machine status bar where the date and time the programmed activities and a...

Страница 17: ...k with relative level Temperature of the side thermal units Temperature of the work plate Status of the Peltier cell On off WARNING The temperatures are visible only when the machine is heating up The...

Страница 18: ...ica it claims bio optica it 18 Publication date 02 03 2020 Rev 00 5 2 SETTINGS MENU The settings menu allows adjusting the machine s operating conditions as well as allowing a log of operating warning...

Страница 19: ...kday Note To correctly set the weekly switch on and off programming see corresponding chapter first set the correct time and date 1 Weekday selection select the current weekday 2 Set the current time...

Страница 20: ...kground no work session has been set The machine will remain in stand by 2 Details of the work sessions set it is possible to display the times of the various work sessions By clicking on the day box...

Страница 21: ...1 PELTIER ON OFF Peltier cell on off enables manually controlling the switching on and off of the desired temperature 2 BOOST enables speeding up the heating phase of the main tank 3 LED work light in...

Страница 22: ...t claims bio optica it 22 Publication date 02 03 2020 Rev 00 5 6 TEMPERATURE SETTINGS In this screen by moving the temperature slider it is possible to adjust the operating temperatures of the units P...

Страница 23: ...ent in the most suitable way for his her work The Operator Panel menu allows enabling or disabling some functions in the instrument This section is accessible to all technicians and allows them to cus...

Страница 24: ...ca Milano S p A Via San Faustino 58 20134 Milan Phone 39 02 21 27 13 1 Fax Italy 39 02 21 53 000 Fax Export 39 02 21 54 155 www bio optica it info bio optica it claims bio optica it 24 Publication dat...

Страница 25: ...only of histological samples fixed and embedded in paraffin Maintenance Periodic annual maintenance must be carried out by qualified personnel authorized by Bio Optica For information please contact...

Страница 26: ...e parafina BEC150 Della classe Of class Der klasse De la classe De la clase Altro dispositivo Other device Sonstiges Produkt Autre dispositif Otro dispositivo Soddisfa tutte le disposizioni della Dire...

Отзывы: