DEFAULT NVRAM (RESET RAM MEMORY)
This function resets the non-volatile RAM (NVRAM) on the Microprocessor PCB. This resets all parameters to the
factory-set default values (i.e., calibration parameters, setpoint parameters, display parameters, etc.).
The controller requires you to enter the Bindicator Service Code (9111) to access this function (unless you have
already entered it during this Manual Mode session), and requests verification before it resets the NVRAM, to
prevent accidental reset of the NVRAM.
96 COPY
The Copy Menu allows you to copy all parameters for the current transducer channel to one or all of the other
transducer channels. This can save you a considerable amount of time if you have multiple channels with the
same or similar parameters.
Follow this procedure to use the Copy Menu.
1. If the controller is in Auto Mode (Auto LED illuminated), press the Scan/Hold Key to put the system in
Manual Mode. The Auto LED turns off.
2. Use the Up Arrow or Down Arrow Keys to scroll to the channel with the parameters you want to copy.
3. Press the Menu Key to display the Main Menu. The display shows:
SonoTracker Series
10 Display
4. Press the up/down arrow keys until the display shows:
SonoTracker Series
90 Service Menu
5. Press the enter key and press the up arrow key to access the Copy Menu. Press Enter. The display shows:
Copy All Params
To All Sensors
6. To copy the parameters to all channels, press the Enter Key and go to Step 7. To copy the parameters to
one channel, press the up arrow Key and press enter, selecting Pick Sensor, go to Step 8.
7. If you selected To All Sensors the display flashed a message acknowledging your selection and returns.
Go to Step 9.
8. YY is the factory-set ID for the channel. If you selected Pick Sensors, the display looks like this:
Sensor:
* YY
Press the Up or Down Arrow Keys to scroll to the channel you want to copy the parameters into. When
you have reached the desired channel, press the Enter Key. The display flashes a message
acknowledging your selection and returns.
9. Press the Esc Key to scroll up the menu tree or press the Scan/Hold Key to return the display to channel
monitoring.
77
www.bindicator.com
SON280020 Rev. C
Содержание SonoTracker Ultrasonic Liquid Level
Страница 1: ...IOM SON280020 Rev C SonoTracker Ultrasonic Liquid Level Installation Operation Manual...
Страница 2: ......
Страница 14: ...10 www bindicator com SON280020 Rev C...
Страница 15: ...11 www bindicator com SON280020 Rev C...
Страница 16: ...12 www bindicator com SON280020 Rev C...
Страница 17: ...13 www bindicator com SON280020 Rev C...
Страница 18: ...14 www bindicator com SON280020 Rev C...
Страница 19: ...15 www bindicator com SON280020 Rev C...
Страница 20: ...16 www bindicator com SON280020 Rev C...
Страница 21: ...venture M E U R E A S T N E M 17 www bindicator com SON280020 Rev C...
Страница 24: ...20 www bindicator com SON280020 Rev C...
Страница 26: ...22 www bindicator com SON280020 Rev C...
Страница 27: ...23 www bindicator com SON280020 Rev C...
Страница 28: ...24 www bindicator com SON280020 Rev C...
Страница 29: ...25 www bindicator com SON280020 Rev C...
Страница 30: ...26 www bindicator com SON280020 Rev C...
Страница 31: ...27 www bindicator com SON280020 Rev C...
Страница 32: ...28 www bindicator com SON280020 Rev C...
Страница 33: ...29 www bindicator com SON280020 Rev C...
Страница 34: ...30 www bindicator com SON280020 Rev C...
Страница 42: ...Figure 10b 30 Relay Set Point Figure 10c 50 Serial PLC 38 www bindicator com SON280020 Rev C...
Страница 43: ...Figure 10d 70 Range Filters Figure 10e 90 Service Menu 39 www bindicator com SON280020 Rev C...
Страница 93: ...89 www bindicator com SON280020 Rev C...
Страница 94: ...90 www bindicator com SON280020 Rev C...
Страница 95: ...91 www bindicator com SON280020 Rev C...