11. Troubleshooting
Fault description
Possible cause
Required measures
Temperature
Set-point temperature is not
reached after specified time.
Chamber door not properly
closed.
Completely close chamber door.
Door gasket defective.
Replace door gasket,
Controller not adjusted.
Calibrate and adjust controller.
Wrong voltage.
Check power supply for voltage of
115V or 230V.
The fan doesn’t turn or turns too
slowly.
Fan speed set too low
Set the fan speed to 100%.
Fan defective.
Contact BINDER service.
Chamber heating permanently,
set-point not held.
Controller defective.
Contact BINDER service.
Pt 100 sensor defective.
Semiconductor relay defective
Controller not adjusted.
Calibrate and adjust controller.
Chamber doesn’t heat up.
Red heating control light in the
display is lit.
Heating element defective.
Contact BINDER service.
Semiconductor relay defective.
Chamber doesn’t heat up.
Red heating control light in the
display is not lit.
Controller display working.
Timer run off.
Program the timer or change to time
function Continuous operation
(chap. 6.3).
Semiconductor relay defective.
Contact BINDER service.
Controller defective.
Chamber without function, only
the green “stand-by” LED is lit
Chamber in stand-by mode.
Press down the ON/OFF button (5)
until the display lights up.
Temperature inside the chamber
too high, Red alarm pilot lamp of
safety device (7a) is lit.
Safety device class 3.1 has
responded.
Check the settings of the
temperature set-point and of the
safety device class 3.1 (chap. 7).
Chamber without any function.
No power supply.
Check connection to power supply.
Chamber fuse has responded.
Check chamber fuse and replace it if
appropriate. If it responds again,
contact BINDER service.
Controller defective.
Contact BINDER service.
Deviations from the indicated
heating-up times.
Chamber fully loaded.
Charge the oven less or consider
longer heating-up times.
Controller
Message „1999” in the controller
display.
Sensor rupture between
sensor and controller.
Contact BINDER service.
The controller returns to Normal
Display from any level.
No button was hit for more
than approx. 30 sec.
Repeat entries, enter the values
rapidly.
Only qualified service personnel authorized by BINDER must perform repair.
Repaired chambers must comply with the BINDER quality standards.
Содержание BF 115
Страница 45: ...BF E1 05 2016 page 45 58 13 Certificates and declarations of conformity 13 1 EU Declaration of Conformity...
Страница 46: ...BF E1 05 2016 page 46 58...
Страница 48: ...BF E1 05 2016 page 48 58...
Страница 49: ...BF E1 05 2016 page 49 58...
Страница 50: ...BF E1 05 2016 page 50 58...
Страница 51: ...BF E1 05 2016 page 51 58...
Страница 52: ...BF E1 05 2016 page 52 58 14 Product registration...
Страница 59: ......