BF (E1) 05/2016
page 28/58
7.
Temperature safety device class 3.1 (DIN 12880)
The temperature safety device class 3.1 serves to protect the chamber, its environment and the charging
material from exceeding the maximum permissible temperature.
Please observe the DGUV guidelines 213-850 on safe working in laboratories (formerly BGI/GUV-I 850-0,
BGR/GUV-R 120 or ZH 1/119, issued by the employers’ liability insurance association) (for Germany).
Figure 6: Temperature safety device class 3.1
Function:
The temperature safety device is functionally and electrically independent of the temperature control
system and if an error occurs it performs a regulatory function.
If you turn the control knob (7) to its end-stop (position 10), the safety device class 3.1 protects the
chamber. If you set it to a temperature a little above the controller’s set-point temperature, it protects the
charging material.
If the safety device has taken over control (identifiable by the red alarm lamp (7a) lighting up and, in case
of the option audible alarm with activated buzzer (chap. 8.1), by the buzzer sounding), proceed as follows:
•
Disconnect the chamber from the power supply.
•
Have an expert examine and rectify the cause of the fault.
•
Restart the chamber (see chap. 5).
Adjustment:
In order to check at which temperature the safety device class 3.1 responds, turn on the chamber and set
the desired set-point on the temperature controller
The sections of the scale from 1 to 10 correspond to the temperature range from 0 °C /
32 °F
to 120 °C /
248 °F
and serve as a setting aid.
1.
Turn the control knob (7) of the safety device with a coin to its end-
stop (chamber protection).
2.
When the set point is reached, turn the control knob (7) to its trip
point (turn it counter-clockwise)
3.
The trip point is identifiable by the red alarm lamp (7a) lighting up.
With the option audible alarm and the buzzer activated (chap. 8.1),
the buzzer sounds as an additional signal. You can turn it off with
switch (10).
4.
The optimum setting for the safety thermostat class 3.1 is obtained
by turning the control knob clockwise by approximately one scale
division, which leads to extinguish the red alarm lamp (7a).
Check the setting regularly and adjust it following any changes of the set-point.
(7)
(7a)
Содержание BF 115
Страница 45: ...BF E1 05 2016 page 45 58 13 Certificates and declarations of conformity 13 1 EU Declaration of Conformity...
Страница 46: ...BF E1 05 2016 page 46 58...
Страница 48: ...BF E1 05 2016 page 48 58...
Страница 49: ...BF E1 05 2016 page 49 58...
Страница 50: ...BF E1 05 2016 page 50 58...
Страница 51: ...BF E1 05 2016 page 51 58...
Страница 52: ...BF E1 05 2016 page 52 58 14 Product registration...
Страница 59: ......