3.4 Location of installation and ambient conditions
Set up the chamber on a flat, even surface, free from vibration and in a well-ventilated, dry location. Align
the chamber using a spirit level. The site of installation must be capable of supporting the chamber’s
weight (see technical data, chap. 12.4). The chambers are designed for setting up inside a building
(indoor use).
CAUTION
Danger of overheating.
Damage to the chamber.
∅
Do NOT set up the chamber in non-ventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
•
Permissible ambient temperature range during operation: +18 °C up to +40 °C /
64.4 °F up to 104 °F
.
At elevated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient
temperature of +22 °C
±
3 °C /
71.6 °F ± 5.4 °F
to which the specified technical data relate.
For other ambient conditions, deviations from the indicated data are possible.
•
Permissible ambient humidity: 70 % r.H. max., non-condensing.
•
Installation height: max. 3000 m /
9842 ft.
above sea level.
When placing several chambers of the same size side by side, maintain a minimum distance of 160 mm
between each chamber. Wall distances: rear 100 mm /
3.94 in
, sides 160 mm /
6.30 in
. Spacing above
the chamber of at least 100 mm /
3.94 in
must also be accounted for.
Two chambers up to size 115 can be stacked on top of each other. For this purpose, place rubber pads
under all four feet of the upper chamber to prevent the device from slipping.
CAUTION
Sliding or tilting of the upper chamber.
Damage to the chambers.
∅
When stacking, place rubber pads under all four feet of the upper chamber.
To completely separate the chamber from the power supply, you must disconnect the power plug. Install
the chamber in a way that the power plug is easily accessible and can be easily pulled in case of danger.
For the user there is no risk of temporary overvoltages in the sense of EN 61010-1:2010.
Do not install or operate the chamber in hazardous locations.
DANGER
Explosion hazard.
Danger of death.
∅
Do NOT operate the chamber in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the vicinity of the chamber.
Содержание BF 115
Страница 45: ...BF E1 05 2016 page 45 58 13 Certificates and declarations of conformity 13 1 EU Declaration of Conformity...
Страница 46: ...BF E1 05 2016 page 46 58...
Страница 48: ...BF E1 05 2016 page 48 58...
Страница 49: ...BF E1 05 2016 page 49 58...
Страница 50: ...BF E1 05 2016 page 50 58...
Страница 51: ...BF E1 05 2016 page 51 58...
Страница 52: ...BF E1 05 2016 page 52 58 14 Product registration...
Страница 59: ......