Program entry as set-point step (example)
W/°C
t/min.
01 02 03 04 05 06 07 08
0
50
100
150
200
250
30 120
180
380
Tolerance
minimum
Tolerance
maximum
On
Off
t/min.
Operation line 1 = air flap
Program table corresponding to the diagram above:
Program
section
Set-point
temp.
Fan Section
time
Operation
line1
Target
section
No. of
cycles
Min.
tolerance
Max.
tolerance
01 50
100
%
00:30:00
Off 1 0
-1999
+9999
02 50
100
%
00:00:01
Off 1 0
-1999
+9999
03 150
100
%
01:30:00
On 1 0 -5 +5
04 150
100
%
00:00:01
On 1 0 -1999
+9999
05 250
100
%
01:00:00
On 1 0 -2 +2
06 250
100
%
00:00:01
On 1 0 -1999
+9999
07 100
100
%
03:20:00
Off 1 0 -1999
+9999
08 100
100
%
00:00:01
Off 1 0 -1999
+9999
Enter now the values of the above program table into the program table of one of the 25 program places
of the program controller MB1:
PGM
09:17:15 15.12.05
For rapid transition phases, do NOT program any tolerance limits in order to allow maximum
heating speed.