CB-S / CB-S-UL (E7) 06/2020
page 29/119
The CO
2
sensor is temperature resistant up to a maximum temperature of 60 °C / 140 °F.
NOTICE
Danger of damage to the CO
2
sensor by excess temperature.
Damage to the CO
2
sensor.
∅
Do NOT autoclave the CO
2
Sensor.
∅
Do NOT expose the CO
2
sensor to hot-air sterilization..
The CO
2
sensor head was especially adjusted for the specific chamber. To avoid confusion, an adhesive
label with a serial number is adhered to the sensor head. When exchanging the sensor, you must repeat
the CO
2
adjustment.
NOTICE
Danger of confusing CO
2
sensors.
Invalid calibration.
Do NOT change the CO
2
sensor head.
Note down the serial number of the CO
2
sensor.
Avoid strong shocks when handling the CO
2
sensor.
NOTICE
Danger of damage to the CO
2
sensor by shocks.
Damage to the CO
2
sensor.
Avoid strong shocks of the CO
2
sensor (by putting it down hard, or dropping).
4.3
Water pan
The water pan permits high humidity without condensation on the inner walls of the CO
2
incubator.
Place the water pan in central position on the bottom of the inner chamber. Place the water pan with its
narrow side
centrally between the front
cams on the bottom of the inner chamber.
Figure 11: Position of the water pan
centrally between the front
cams (arrows)
Содержание 9040-0189
Страница 109: ...CB S CB S UL E7 06 2020 page 109 119 26 9 Dimensions CB S CB S UL 170 Dimensions in mm...
Страница 110: ...CB S CB S UL E7 06 2020 page 110 119 CB S CB S UL 260 Dimensions in mm...
Страница 111: ...CB S CB S UL E7 06 2020 page 111 119 27 Certificates and declarations of conformity 27 1 EU Declaration of Conformity...
Страница 112: ...CB S CB S UL E7 06 2020 page 112 119...
Страница 113: ...CB S CB S UL E7 06 2020 page 113 119 28 Product registration...