97
DD1050
10 CUSTOMISING METROLOGICAL PARAMETERS
Press the calibration button to access the metrological parameter customising options (parameters
linked more sensibly to the scale's legal operating mode).
I--(Figura12-1.jpg)--I
figure 10.1 - CPU board disassembly
Key:
1. Adhesive label that seals the hole that accesses the calibration switch
2. Calibration switch
Proceed in the following way, with reference to Figure 10.1:
• Switch on the terminal.
• Remove the adhesive label (ref. 1) that seals the hole that accesses the calibration switch.
• Operate the calibration switch (ref.2) using a non-metal tool.
CAUTION
It will be impossible to use the weighing terminal for commercial transactions with third parties
if the metrological seals are removed. Contact the Assistance Network's After Sales Service to
have them replaced.
1
2