Instructions for use

BW 585

Blood and Infusion Warmer

0123

EN - Edition 03 / 2007

Содержание BW 585

Страница 1: ...Instructions for use BW 585 Blood and Infusion Warmer 0123 EN Edition 03 2007...

Страница 2: ...ork has been carried out BIEGLER Instructions for use BW 585 14 TECHNICAL DATA Device Blood and infusion warmer Type designation BW 585 Voltage 230 V 50 60 Hz or 110 V 50 Hz check the voltage specifie...

Страница 3: ...C Values higher or lower than the ranges specified above may cause damage to the device or its accessories 16 EN Edition 03 2007 BIEGLER Instructions for use BW 585 Table of contents 1 Warnings and s...

Страница 4: ...function deviating from the instructions for use do not use the device and return it to the service centre The periodic technical safety inspections must be carried out according to the section perio...

Страница 5: ...ns Electromagnetic is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones an...

Страница 6: ...w The customer or the user of the BW 585 should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic dis...

Страница 7: ...ns for use before putting the system into operation 3 1 SETTING UP PROCEDURE Fix the BIEGLER BW 585 firmly to the infusion stand using the clamps at the back Diag 1 9 Only use infusion stands or poles...

Страница 8: ...is switched off To reset the device or switch off the alarm the device must first be disconnected from the power supply Attention The temperature alarm can also be set off externally e g by exposure t...

Страница 9: ...The manufacturer s warranty will be void if it is found that the maintenance disinfection and inspection instructions have not been followed according to the instructions for use the device was damag...

Страница 10: ...all the alarms are active but the heating is switched off The BW 585 now slowly cools down When the temperature drops below the 36 5 C threshold the low temperature alarm should be active For reasons...