Page 8 of 19 M7998 Issue 2.3
5. INSTALLATION.
5.1.
Electrical Safety and Installation.
5.1.1. The Integrity Reaction Station and equipment is designed to be used safely under the following
conditions:-.
•
Indoor use only.
•
Altitude up to 2000 meters.
•
Temperatures between 5
°
C and 40
°
C.
•
Maximum relative humidity 80% for temperature up to 31
°
C decreasing linearly to 50% relative
humidity at 40
°
C
•
Mains supply voltage fluctuations up to
±
10% of the nominal voltage.
•
Transient overvoltage’s typically present on the mains supply. (Overvoltage category II).
•
Applicable rated pollution degree 2.
5.2.
This equipment must be earthed / ground to a fixed earth / grounded mains socket outlet. The
mains supply is to be earthed / grounded in accordance with current and local legislation.
5.3.
Ensure only the correct rated mains input fuses are fitted. (Where applicable ensure the correct
mains cord and moulded IEC plug and lead set fuse, if fitted). See Technical Specification Section 8
of this instruction book.
5.4.
It is recommended this product be connected to a mains supply source which incorporates an RCD
or GFCI device.
5.5.
Do not install this product or accessories on a surface which may become wet.
5.6.
Mains Supply.
5.6.1. Check the voltage on each product data label on your Reaction Station and those of any
accompanying electrical accessory. Ensure the rating conforms to your local supply.
5.6.2. The reaction station is supplied with a Mains cord and moulded plug wired as follows.
Green / Yellow
or
Green
=
Earth / Ground
Blue
or
White
=
Neutral
Brown
or
Black
=
Live / line hot.
5.6.3. Solubility IR box is equipped with one universal power supply to be plugged into an electrical outlet
using the lead supplied suitable for the country of use.
5.7.
Coolant Supply.
5.7.1. It is recommended the minimum coolant flow rate be 2.5 litres/min at 5
°
C. The coolant system
should be able to dissipate heat energy at 1000W. Running all cell positions at -30
°
without the
recommended coolant supply may cause stress on the electrical components, resulting in reduced