Page 10 of 19 M7998 Issue 2.3
Note: The PC should be set up and located near the Clarity Station.
5.10.
Installing Clarity IR box
5.10.1. Place the Multi-Channel IR Box onto a clean, flat, well-ventilated work area on the left side of the
Clarity Reaction Station, ensuring there are no combustible materials in the vicinity.
5.10.2. Attach the RS232 cable to the RS232 port on the clarity IR box.
5.10.3. Attach other end of the RS232 cable to the RS232 port on the Integrity 10 reaction station as
illustrated. Plug the TNX125 to the appropriate electrical outlet and then to Clarity IR Box.
5.10.4. Install IR probes to Multi-IR Box. For turbidity probes (ATS10230) match launch with launch and
return with the return. For Shoot-Trough probes (ATS10360/1), the connections are reversed.