Bibby Sterilin Electrothermal ATS10232E Скачать руководство пользователя страница 1

Page 1 of 19                                                                                                                                     M7998 Issue 2.3 

 

 
 
 
 
 
 
 

 

 
 
 
 
 
 

CLARITY STATION 
 

ATS10232E 
 

Multi-Input Light Measurement Module. 
 
 
 
INSTRUCTION BOOK. 

 
 

Содержание Electrothermal ATS10232E

Страница 1: ...Page 1 of 19 M7998 Issue 2 3 CLARITY STATION ATS10232E Multi Input Light Measurement Module INSTRUCTION BOOK...

Страница 2: ...n Certificate Page 19 The copyright of this instruction book is the property of Bibby Scientific Limited This instruction book is supplied by Bibby Scientific Limited on the express understanding that...

Страница 3: ...ve way of determining the solubility of a compound by monitoring the turbidity of the mixture The complete system for solubility consists of the following Multi Channel Infra Red box ATS10232E Integri...

Страница 4: ...one s self or other people The Responsible Body and the Operator should read and be familiar with this Instruction book in order to preserve the protection afforded by the equipment To prevent injury...

Страница 5: ...being heated and ensure all hazards e g explosion implosion or the release of toxic or flammable gases that might arise have been suitably addressed before proceeding When heating certain substances t...

Страница 6: ...Section 10 Neither the Reaction Station or Clarity System are spark flame or explosion proof and no part of the equipment has not been designed for use in hazardous areas in terms of BSEN 60079 14 199...

Страница 7: ...m Description Qty 1 IR control box ATS10232E 1 2 IR probe ATS10230 Not Included 0 3 Instruction Book M7998 1 4 PSU mains leads illustration may be different to that shown depending on destination A R...

Страница 8: ...itted See Technical Specification Section 8 of this instruction book 5 4 It is recommended this product be connected to a mains supply source which incorporates an RCD or GFCI device 5 5 Do not instal...

Страница 9: ...disk drive Mouse or similar pointing device 5 8 2 RAM and Hard Disk Space 32 MB of RAM 128 MB recommended 20 MB free hard disk space 5 8 3 Supported Operating Systems The PC control software is compat...

Страница 10: ...ustible materials in the vicinity 5 10 2 Attach the RS232 cable to the RS232 port on the clarity IR box 5 10 3 Attach other end of the RS232 cable to the RS232 port on the Integrity 10 reaction statio...

Страница 11: ...7 10 5 11 1 Attach RS232 cable to the RS232 port on the back side of the Multi Temp Box USING Null Modem Adaptor 5 11 2 Attach the other side of the RS232 cable to the Aux GSIOC Port on the Clarity IR...

Страница 12: ...Page 12 of 19 M7998 Issue 2 3...

Страница 13: ...Complete steps 7 1 4 to 7 1 6 for all cells requiring calibration 6 2 Shoot Through Probe Calibration 6 2 1 Open Calibration screen from the Tools drop down menu 6 2 2 Adjust the variable in the calib...

Страница 14: ...rbidity probes into each vessel 6 3 4 In the Integrity PC software use the Set Plateau Profile Template to setup the test method of each well according to is percentage weight For example if the weigh...

Страница 15: ...Resolution 0 1 Power Supply Input Voltage 10 to 15V d c or 7 to 11V a c Connector 2 1mm power in connector Power consumption 10W Interfaces RS232 C via DB9F connector Aux RS232 C RS485 GSIOC via DB9M...

Страница 16: ...or IR box and probes 8 4 Spillage and Decontamination If the equipment has been exposed to spillage or contamination the Responsible Body is responsible for carrying out appropriate decontamination If...

Страница 17: ...er information For end users outside of the EU consult applicable regulations 9 4 This product should only be dismantled for recycling by an authorised recycling company This product and accessories m...

Страница 18: ...ain Tel 44 0 1785 812121 Fax 44 0 1785 810405 General enquiries info bibby scientific com Order enquiries sales bibby scientific com Technical support electrothermalhelp bibby scientific com www elect...

Страница 19: ...leting YES NO as applicable and by providing details in section 2 below A Blood body fluids Pathological specimens YES NO Provide details if YES B Biodegradable material that could become a hazard YES...

Отзывы: