Instruction for Use
27
The stickers are available from BHM.
Only stickers from BHM may be used.
Before programming
There must not be a battery in the
device during programming!
Before programming, please remove the
socket cover and socket seal (Symbolic
image 17). To do this, please use
the BHM screwdriver supplied and
push downward through the loop
onto the cov er. This is then easy to
fold away. Then simply remove the
socket seal.
that the headband or softband fits comfor-
tably on the patient. The conductor plate
of the integrated adapter must be placed
behind the ear on the mastoid. If necessa-
ry, the position can be varied accordingly.
The patient must be instructed accordingly.
Wearing the headband or softband should
be perceived as comfortable as possible.
Adhesive adapter
Adhesive adapters (adhesive pad plus
sticker) for the left and right wearing side
are included with the hearing system. In-
structions for using the adhesive adapters
can be found in this instructions for use.
Pay attention to the correct positioning of
the adhesive adapter behind the ear on the
patient's mastoid. The skin area where the
adhesive adapter is to be attached must
be clean, dry and free of hair. If necessary,
the position can be varied accordingly. The
patient must be instructed accordingly.
Symbolic image 17