8
Notice
- The default value is set to “0.5”
- When the air pressure is not between “0.3” – “1.0”. “Eb will be shown on the display and the value
will not be memorized.
6) After setting the lower limit, press the Gear Ration Select Key. The indicator will change to “C2”
(Upper air pressure limit setting).
7) By pressing the FWD/REV rotation key, the indicator will show “0.0” and “Ec”.
8) By stepping on the foot pedal of the delivery system. The indicator will show the current air pres
-
sure.
Notice
The displayed air pressure is only a guideline. Measure the correct air pressure with the air meter.
9) Maintain air pressure between “2.0” – “4.0”. Press the P2 key until it beeps (more than 3 seconds)
to memorize the upper limit of the air pressure.
Notice
The Default value is set to “3.0”.
When the air pressure is not between “2.0” – “4.0” the display will show “Ec”, and the value will not be
memorized.
10) Press both the P2 key and the Speed Adjustment Key (+) at the same time until you hear a beep
(more than 3 seconds). When you hear the beep the setting is complete.
6-2 Light Brightness Adjustment
(unit of Display = V)
Brightness of the Fiber Optic can be adjusted with this function.
1) Press the P2 Key and Speed Adjustment Key (+) at the same time until its beeps. (more than 3
seconds).
2) After the beep, the indicator will change to “C1”. Pressing the Gear Ration Select Key will change
the indicator and the Gear Ration Lamp (C1 ->C2 ->C3-> C1…) Set the indicator in C3
(Light Brightness Adjustment Mode). Press FWD/REV Rotation Key to proceed.
3) When The indicator its default value (“3.5”), Press the Speed Adjustment Key (+/-) to adjust the
brightness.