background image

3

 Caution

Unit

- Grasp cord by the plug to remove from the outlet. Do not pull or yank the cord itself. It may cause 

the wire to malfunction. 

- Care should be taken not to place the AC cord near a gas burner. Never attempt to repair a burned 

motor cord. Always replace it with a new cord.

- Prior to use, always check for vibration, noise and overheating. If any abnormalities are detected, 

stop using immediately and contact your dealer.

- Be careful not to spill water onto the unit as it may result in a fire or an electric shock due to a short 

circuit. 

Motor, Hand piece 

- Do not use this product under strong stress for extended periods of time. It can cause overheating.

- Do not connect or disconnect the cord until the drive motor has completely stopped.

- Do not connect/disconnect the hand piece during operation.

- Verify that the Speed Control Switch is adjusted within the allowable speed before use.

- Connect only 4-hole, 5-hole or 6-pin tubing.

- Air Requirements: Dry, free from contamination and oil. Use a compressor with a dry air system/ 

Install an air filter if necessary/ Blow out the lines before installation.

- Do not autoclave (or any other high temperature sterilization) Control Box, AC Adaptor, Motor Code.

- The user shall be responsible for operation and maintenance and inspection.

 Notice

During rotation, the motor and motor cord may affect computer and LAN cables. Noise could be heard 

during operation near radio receivers. 

After treatment, immediately turn the power and air supply off. Remove the power cord if the unit is 

not to be used for a long time.

1. Specifications

Model

Maxso E600P

Speed

2,000 ~ 40,000 rpm

Torque

3.4 Ncm

Rated Input

AC 24V / DC36

Noise Level

<60dB

Dimensions

W127 x D149 x H54 (mm)

Содержание Maxso E600

Страница 1: ...Electric Micromotor System Instruction for Use Maxso E600...

Страница 2: ...r by holding on to the base of the plug and not the cord To Prevent Electric Shock Use an electrical outlet which is grounded If you feel any abnormality such as excessive vibration heat or noise stop...

Страница 3: ...extended periods of time It can cause overheating Do not connect or disconnect the cord until the drive motor has completely stopped Do not connect disconnect the hand piece during operation Verify th...

Страница 4: ...Gear Ration Select Key 5 FWD REV Select Key 6 Program 1 Button 7 Program 2 Button 8 Speed Adjusters 1 9 10 2 3 4 6 5 7 8 12 11 13 14 16 15 9 Motor 10 Motor Cord 11 Power Switch 12 AC Adapter Connecto...

Страница 5: ...urbine hose connector and tighten completely Caution Do not use any other hose than the ones provided with the unit Do not use contaminated air Blow out the lines before installation Caution Never use...

Страница 6: ...requires lubrication E Type micro motor Following lubrication stand the hand piece attachment vertically until the excess solution drains Wipe the hand piece then attach it to the motor Do not connec...

Страница 7: ...system is not high enough by setting the upper and lower limit of the air pressure Notice Do this adjustment on initial set up of the unit Once it is adjusted it will be memorized and you won t need t...

Страница 8: ...imit of the air pressure Notice The Default value is set to 3 0 When the air pressure is not between 2 0 4 0 the display will show Ec and the value will not be memorized 10 Press both the P2 key and t...

Страница 9: ...lave sterilization method Sterilization is required the first time you use the unit as well as after each patient as noted below Autoclaving 1 Turn off the power 2 Detach the motor from the motor cord...

Страница 10: ...internal cir cuit more than 3 7V Contact your dealer E8 Overheating error High temperature safety system activated due to extended use Wait for a while and let it cool down before using again 8 Maint...

Страница 11: ...r limit of air pressure Check the air pressure on the delivery system Do Foot Air calibration Beeps when turning on the switch You are stepping on the foot pedal when turning on the power switch Do no...

Страница 12: ...owing 1 Defects damage or other conditions caused in whole or in part by misuse abuse negligence alteration accident freight damage negligent storage tampering or failure to seek and obtain repair or...

Страница 13: ...horized to sell the product at issue within the medical animal health or dental market as the case may be in which purchaser intends to use the product 11 4 Exclusive Remedy Consequential Damages Disc...

Страница 14: ...English 31626669008 Dutch 0197 Printed in Canada File ENI019R1 Federal law restricts this device to sale by or on the order of a dentist physician or any other practitioner licensed by the law of the...

Отзывы: