Electric Micromotor System

Instruction for Use

Maxso

®

 E600

Содержание Maxso E600

Страница 1: ...Electric Micromotor System Instruction for Use Maxso E600...

Страница 2: ...r by holding on to the base of the plug and not the cord To Prevent Electric Shock Use an electrical outlet which is grounded If you feel any abnormality such as excessive vibration heat or noise stop...

Страница 3: ...extended periods of time It can cause overheating Do not connect or disconnect the cord until the drive motor has completely stopped Do not connect disconnect the hand piece during operation Verify th...

Страница 4: ...Gear Ration Select Key 5 FWD REV Select Key 6 Program 1 Button 7 Program 2 Button 8 Speed Adjusters 1 9 10 2 3 4 6 5 7 8 12 11 13 14 16 15 9 Motor 10 Motor Cord 11 Power Switch 12 AC Adapter Connecto...

Страница 5: ...urbine hose connector and tighten completely Caution Do not use any other hose than the ones provided with the unit Do not use contaminated air Blow out the lines before installation Caution Never use...

Страница 6: ...requires lubrication E Type micro motor Following lubrication stand the hand piece attachment vertically until the excess solution drains Wipe the hand piece then attach it to the motor Do not connec...

Страница 7: ...system is not high enough by setting the upper and lower limit of the air pressure Notice Do this adjustment on initial set up of the unit Once it is adjusted it will be memorized and you won t need t...

Страница 8: ...imit of the air pressure Notice The Default value is set to 3 0 When the air pressure is not between 2 0 4 0 the display will show Ec and the value will not be memorized 10 Press both the P2 key and t...

Страница 9: ...lave sterilization method Sterilization is required the first time you use the unit as well as after each patient as noted below Autoclaving 1 Turn off the power 2 Detach the motor from the motor cord...

Страница 10: ...internal cir cuit more than 3 7V Contact your dealer E8 Overheating error High temperature safety system activated due to extended use Wait for a while and let it cool down before using again 8 Maint...

Страница 11: ...r limit of air pressure Check the air pressure on the delivery system Do Foot Air calibration Beeps when turning on the switch You are stepping on the foot pedal when turning on the power switch Do no...

Страница 12: ...owing 1 Defects damage or other conditions caused in whole or in part by misuse abuse negligence alteration accident freight damage negligent storage tampering or failure to seek and obtain repair or...

Страница 13: ...horized to sell the product at issue within the medical animal health or dental market as the case may be in which purchaser intends to use the product 11 4 Exclusive Remedy Consequential Damages Disc...

Страница 14: ...English 31626669008 Dutch 0197 Printed in Canada File ENI019R1 Federal law restricts this device to sale by or on the order of a dentist physician or any other practitioner licensed by the law of the...