7
Picture 5 (a)
2.1.2 Testing interface introduction
Picture 5(a) shows root canal file in the dynamic
operation of teeth in the solution. When not
examined, no root canal file; after detected data,
the corresponding schematic diagram will display
according to the different position in the file teeth;
When the tip of the root canal file arrived near the root
of tooth, tooth solution of plane map will display the
blue circle represents the position of amplification.
Picture 5(b) is enlarged display area. When
the root canal file did not reach near the root tip, the
entire root apex area is gray. When the needle slowly
close to the root tip, color scale will gradually fill, the
alarm sound will be more rapid(non-silence state);
When the needle reached the preset of the root file,
the alarm will ring often to remind that has reached
the apex; When file tip needle penetration, red dot
prompt is displayed below the enlarged area with
accompanying alarm.
Picture 5(c) is the position of the apical region
of the root, The position of red arrows point is preset
root tip, When the red arrow is located at a scale (the
red arrows are located in the 0 scale for product
factory), the measurement results the measurement
results to or beyond the scale, will display ”APEX”
with alarm. The method for adjusting the position of
the red arrow is long press the volume key for 1
second, and the moving range is 0 ~ 0.4.
Picture 5 (b)
Picture 5 (c)