11.2 Guidance and manufacturer statement - Electromagnetic Immunity
Guidance and manufacturer statement - Electromagnetic Immunity
The ApexPilot is intended for use in the electromagnetic environment specified below. The
customer or the user of the ApexPilot shouId assure that is used in such an environment.
Immunity test
IEC60601 test leve
Compliance Leve
Electromagnetic
environment - guide
Electrostatic
discharge(ESD)
EN 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity shouId
be at least 30%.
Electrical fast
transient/burst,
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
Output lines
± 2 kV for power
supply lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge IEC
61000-4-5
±1 kV differential
mode
±1 kV differentia
mode
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5 % UT
( > 95 % dip in UT)
For 0.5 cycles
40% UT
( 60 % dip in UT) for
5 cycles
< 5 % UT 70 % UT
( 30 % dip in UT)
for 25 cycles
< 5 % UT
( > 95 % dip in
UT) for
5 s
< 5 % UT
( > 95 % dip in UT)
For 0.5 cycles
40% UT
( 60 % dip in UT) for
5 cycles
< 5 % UT 70 % UT
( 30 % dip in UT)
for 25 cycles
< 5 % UT
( > 95 % dip in
UT) for
5 s
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the ApexPilot requires
continued operation during
power mains interruptions,
it is recommended that the
ApexPilot be powered
from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.