INTEGRATED ELECTRICAL CABINET

 

Installation, Operation, and Service Manual

 

 

Copyright

 2016:  

Bevco Engineering Company, Inc., 

W222N5739 Miller Way,  Sussex, WI 53089 USA 

IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems 

                        Page 3 of 14 

1.0  DESCRIPTION OF OPERATION 

 

This panel incorporates several features desirable for system installations to minimize down time, 
protect the X-Ray Generator electronics, and to reduce operational delays after a power outage.  This 
product replaces the power panel, the lighting control relay box, and the Emergency Power Off 
controls.  The panel has a main circuit breaker, Q1, provides 

fully integrated “X-Ray ON” warning light 

control and a relay to reduce the room lighting during the procedure.  When the main circuit breaker is 
turned off, all power circuits within the panel will be de-energized.     
 
Q1 provides the disconnect means and lock-out and tag-out (LOTO) the X-Ray Generator power circuit 
for maintenance purposes.  The K2 contactor will open with any loss of power or by pressing any 
Emergency Power Off (EPO) pushbutton.  The contactor control circuit is factory configured to 
automatically re-energize the X-Ray Generator upon restoration of facilities power.   The control circuit 
may be re-configured to require the operator to manually restart the equipment once the incoming 
power has been restored.  This protects the sensitive electronic circuits of the X-Ray Generator from 
sags and surges that immediately follow power loss from blackouts, storms, utility reclosure operations, 
and out of phase automatic transfer switch operations. 
 
The SC1 cabinet is protected by an electronic circuit breaker, Q3, and controlled by contactor K6. The 
contactor will open with any loss of power or by pressing any Emergency Power Off (EPO) pushbutton. 
 
The control circuits for the EPOs are low voltage 24 VDC and are fully powered from within the panel.  
The restart functionality and EPO circuitry is controlled with a safety relay, K10.  See page 11 for detail. 
 
The white SAFETIES OK indicator light on the front of the panel is illuminated when none of the EPOs 
are pressed.  When the white light is active, pressing the green START pushbutton will cause the X-
Ray Generator to be energized.  The green START button will illuminate, and the white SAFETIES OK 
light will go off.  Pressing the STOP button will de-energize the system.  Any EPO pressed while the 
system is energized will result in the immediate de-energizing of the X-Ray Generator system. 
 
If an EPO is pressed at any time, the EPO must be reset which will cause the SAFETIES OK light to 
activate.  Then the START button will activate the system. 
 
The EPOs provided with the IEC include additional Normally Closed contacts for connections to a UPS 
system.

 

 
 

IMPORTANT: 

If building power is removed from the panel while the X-Ray system is energized, the power to the X-
Ray system will be restored when building power returns without any human interaction.  The X-Ray 
system can then be restarted normally. 
 
Additional provisions are made to integrate the “X-Ray ON” warning lights and room lighting with the X-
Ray Equipment.  The facility lighting panel provides 120- or 277-volt power that is controlled by 
contacts relays of K4 and K5 mounted in the IEC.  The signal controlling the relays comes directly from 
the Siemens Generator/Power Cabinet.  The IEC will accept signals from the generator as either Dry 
Contacts or 24 Volts AC or DC.  Note the maximum connected load to the X-Ray On lights is 20 Amps. 

 
 

Содержание IECAX480V125A

Страница 1: ...ications Installation Operation and Service Manual SIEMENS HEALTHCARE INTEGRATED ELECTRICAL CABINET for ARTIS Q Q ZEN ONE ARTIS ZEEGO ZEEGO ZEN ARTIS ZEE MULTIPURPOSE SINGLE PLANE IECAX480V125A Main Disconnect Panel ...

Страница 2: ... of Bevco Engineering Company Troubleshooting and servicing should be performed by a qualified electrician De energize and lockout the main disconnect before servicing this panel Table of Contents 1 0 DESCRIPTION OF OPERATION 3 2 0 SYSTEM CONFIGURATION 4 3 0 SPECIFICATIONS 5 ELECTRICAL SPECIFICATIONS 5 SEISMIC SPECIFICATIONS 5 ENVIORNMENTAL SPECIFICATIONS 5 ELECTRICAL WIRING DIAGRAM 5 4 0 PANEL DI...

Страница 3: ...3 and controlled by contactor K6 The contactor will open with any loss of power or by pressing any Emergency Power Off EPO pushbutton The control circuits for the EPOs are low voltage 24 VDC and are fully powered from within the panel The restart functionality and EPO circuitry is controlled with a safety relay K10 See page 11 for detail The white SAFETIES OK indicator light on the front of the pa...

Страница 4: ...R K4 ROOM LIGHT RELAY K5 WARNING LIGHT RELAY EMERGENCY POWER OFF CIRCUITS EPO1 EPO2 LIGHTING POWER PANEL 120 277V KEY TO COMPONENTS Q1 MAIN CIRCUIT BREAKER Q2 GENERATOR 2 CIRCUIT BREAKER Q3 GENERATOR 1 CIRCUIT BREAKER F2 POWER SUPPLY FUSES F4 DC CIRCUIT BREAKER G01 DC POWER SUPPLY K1 SAFETY CONTACTOR K2 X RAY GENERATOR CONTACTOR K4 ROOM LIGHT RELAY K5 X RAY ON WARNING LIGHT RELAY K6 SYSTEM CABINET...

Страница 5: ...nd Neutral 125 Amps cULus Short Circuit Current Rating is 35 000 Amperes RMS Symmetrical 480V SEISMIC SPECIFICATIONS This Integrated Electrical Cabinet has been certified by an independent California structural engineer in conformance with the shake testing requirements of ICC AC 156 The California OSHPD number is OSP 0457 10 The seismic performance characteristics are as follows SDS g 2 56 z h 1 ...

Страница 6: ...el may be recessed into the wall up to 5 inches for semi flush installations See Illustrations below for typical front and side views with the mounting hole pattern Refer to the drawing set provided with the panel for total weight and center of gravity dimensions PANEL EXTERNAL PHYSICAL DIMENSIONS Parameter Dimensions Height 30 0 inches 762 mm Width 20 0 inches 508 mm Total Depth 9 44 inches 244 m...

Страница 7: ... CABINET Installation Operation and Service Manual Copyright 2016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 7 of 14 CENTER OF GRAVITY ...

Страница 8: ...INET Installation Operation and Service Manual Copyright 2016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 8 of 14 MOUNTING HOLE DIMENSIONS ...

Страница 9: ...ectly to the load side of contactor K2 terminals 2T1 4T2 and 6T3 5 Connect the System Cabinet SC1 directly to the load side of contactor K6 terminals 2T1 4T2 and 6T3 6 Connect the IEC Emergency Power Off EPO circuits to terminals 1 and 2 of X02 All EPOs should have normally closed contacts wired in series See EPO wiring diagram 7 Room light controls are on terminals X05 1 through 4 Verify first if...

Страница 10: ...e Manual Copyright 2016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 10 of 14 INSTALLATION continued TERMINAL LAYOUT Located in the bottom left of the Disconnect Panel EPO WIRING ...

Страница 11: ...Installation Operation and Service Manual Copyright 2016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 11 of 14 LIGHTING WIRING DOOR SWITCH WIRING ...

Страница 12: ...k the OUT light on the safety relay if not solid green press and hold the RESET button for five seconds then release The lights should flash and the DEVICE and IN light should be green Press the Start button contactor K2 should pull in If the DEVICE light is red the safety relay is defective and must be replaced For other Safety Relay Conditions refer to the diagram and chart next page PDA Off cir...

Страница 13: ...016 Bevco Engineering Company Inc W222N5739 Miller Way Sussex WI 53089 USA IECAX480V125A OM1_R51 Siemens IEC Manual for AX Systems Page 13 of 14 K10 Safety Relay configurations The Safety Relay DIP switches are set as follows from the factory 1 2 3 4 5 6 7 8 On On Off On On On On Off ...

Страница 14: ...ervicing 8 0 REPLACEMENT PARTS 9 0 DOCUMENT REVISION HISTORY REV DATE AUTHOR DESCRIPTION OF CHANGE 1 10 03 2013 RJB Initial Release 2 02 27 2015 RJB Added Artis Tradenames on first page 3 03 09 2015 XYUE Update the manual based on schematic 50 03 30 16 DRO IEC Redesign 51 07 10 18 JT Update of system config data PARTS TECHNICAL SUPPORT AND SERVICE FOR INQUIRIES DURING INSTALLATION CONTACT BEVCO EN...

Отзывы: