For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic emissions
The ME 90 is intended for use in the electromagnetic environment specified below. The customer or the
user of the ME 90 should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The ME 90 uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11
Class B
The ME 90 is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions IEC 61000-3-2
Class A
Voltage fluctuations/ flicker emissions Complies
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
The ME 90 is intended for use in the electromagnetic environment specified below. The customer or the
user of the ME 90 should assure that it is used in such an environment.
IMMUNITY test IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV
contact ±8
kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power sup-
ply lines
±2 kV for power
supply lines
Mains power quality should be that of a typical commer-
cial or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Mains power quality should be that of a typical commer-
cial or hospital environment.