113
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
The BM 48 is intended for use in the electromagnetic environment specified below. The customer or the user of the BM 48 should
assure that it is used in such an environment.
IMMUNITY test
IEC 60601 test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are co-
vered with synthetic material,
the relative humidity should be
at least 30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environ-
ment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.