Beurer BGL40 mg/dL
28
If the results are outside the specified range, check the following possible causes:
Cause
Measure
The code number on the display is not
the same as that of the test strip
container.
Code the monitor (see "6.1 Checking
code number (coding)“ page 24).
•
You did not dispose of the first drop
of control solution
•
The vial tip was not wiped clean.
•
The vial was not shaken sufficiently.
Correct the cause and repeat the
test.
Control solution has expired or is
contaminated.
Repeat the test with a new vial of
control solution.
Control solution or monitor is too warm
or too cold.
Bring control solution and monitor to
room temperature (+20°C to +25°C)
and repeat test.
Damaged test strips
Repeat the test with a new test strip.
Old test strips
Open a new test strip container.
Repeat the test.
A problem with the monitor
Contact our Customer Service
department.
WARNING
If you receive repeated results outside the specified range with the
control solution,
do not continue using the system to measure your
blood glucose level
. Contact our Customer Service department.