Beurer BGL40 mg/dL
27
4
Check the code number. If the numbers are different remove the test strip
and code the monitor (see "6.1 Checking code number (coding)” page
24).
5
Press the 'M' button to switch to control mode. In control mode test results
are not saved and your statistics are not skewed.
6
CTL is shown on the display.
7
Shake the control solution well before use. Unscrew the cap and squeeze
out one drop. Wipe the first drop away and squeeze out another drop.
8
Do not apply the drop directly to the test strip, otherwise the control
solution remaining in the vial tip may be contaminated by contact with the
test strip. First place the drop on your fingertip or on a clean surface.
Then hold the drop to the slot for taking blood on the test strip. The
solution is sucked into the slot. Wipe the tip of the vial with a clean, dry
paper towel.
9
When the slot is filled with solution, the monitor carries out the test. The
monitor counts back for approximately six seconds. The result is shown in
the display.
10
Check that the test result is within the specified range of the control
solution. The result range is printed on the test strip container.
Expected results
At room temperature the test results with the control solution should be within the
range printed on the test strip container for approximately 95% of all tests.
WARNING
The results range printed on the test strip container is only applicable to
the control solution.
It is not a recommended value for your blood
glucose level.