25
T
able 2
Guidance and manufactur
er
’s
declar
ation -
electr
omagnetic Immunity
Immunity T
est
IEC 60601-1-2
Test level
Compliance level
Electr
ostatic
dischar
ge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV
, ±4 kV
, ±8
kV
, ±15 kV air
±8 kV contact
±2 kV
, ±4 kV
, ±8
kV
, ±15 kV air
Electrical fast
tr
ansient/burst
IEC 61000-4-4
Not applicable
Not applicable
Sur
ge
IEC 61000-4-5
Not applicable
Not applicable
V
oltage
dips, short
int
erruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
Not applicable
Not applicable
Immunity T
est
IEC 60601-1-2
Test level
Compliance level
P
ower
fr
equency
magnetic field
IEC 61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Conduct
ed RF
IEC61000-4-6
Not applicable
Not applicable
Radiat
ed RF
IEC61000-4-3
10 V/m
80 MHz – 2,7
GHz
80 % AM at 1
kHz
10 V/m
80 MHz – 2,7
GHz
80 % AM at 1
kHz
NO
TE UT is the a.c. mians voltage prior t
o
application of
the t
est level.