24
EMC Guidance
The ME EQUIPMENT or ME S
Y
S
TEM is suitable for home
healthcar
e envir
onments
W
arning:
Don’t use near active HF sur
gical equipment
and the RF shielded r
oom of an ME syst
em for magnetic
resonance imaging, wher
e the int
ensity of EM disturbances
is high.
W
arning:
Use of this equipment adjacent t
o
or stacked with
other equipment should be avoided because it could r
esult in
impr
oper oper
ation. If such use is necessary, this equipment
and the other equipment should be obser
ved t
o
verify that
they ar
e oper
ating normally
.
W
arning:
Use of accessories, tr
ansducers and cables other
than those specified or pr
ovided by the manufactur
er of
this equipment could r
esult in incr
eased electr
omagnetic
emissions or decr
eased electr
omagnetic immunity of this
equipment and r
esult in impr
oper oper
ation.
”
W
arning:
P
o
rtable RF communications equipment (including
peripher
als such as ant
enna cables and e
x
ternal ant
ennas)
should be used no closer than 30 cm (12 inches) t
o
any
part of the equipment , including cables specified by the
manufactur
er
. Other
wise, degr
adation of the performance of
this equipment could r
esult.
T
echnical description:
1. All necessary instructions for
maintaining BASIC
SAFETY and ESSENTIAL PERFORMANCE with r
egar
d
to
electr
omagnetic disturbances for the e
x
cept
ed
service life.
2. Guidance and manufactur
e
r’
s declar
ation
-electr
omagnetic emissions and Immunity
T
able 1
Guidance and manufactur
er
’s
declar
ation -
electr
omagnetic emissions
Emissions t
est
Compliance
RF emissions, CISPR 11
Gr
oup 1,
RF emissions, CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Not applicable
V
oltage fluctuations/ flicker
emissions, IEC 61000-3-3
Not applicable