background image

Beurer GL50 

37

Concentration of  

tested substances

50–100 mg/dL

(2.8–5.6 mmol/L)

250-350 mg/dL

(13.9-19.4 mmol/L)

Acetaminophen

7 mg/dL

(0.46 mmol/L)

6.6 mg/dL

(0.37 mmol/L)

4.5%

Ascorbic acid

4 mg/dL

(0.23 mmol/L)

3.3 mg/dL

(0.18 mmol/L)

5.1%

Bilirubin

3.3 mg/dL

(0.06 mmol/L)

0.1 mg/dL

(0.01 mmol/L)

-1.4%

Cholesterol

400 mg/dL

(10.34 mmol/L)

-6.8 mg/dL

(-0.38 mmol/L)

-6.2%

Creatinine

30 mg/dL

(2.65 mmol/L)

0.0 mg/dL

(0.00 mmol/L)

-0.1%

Dopamine

2.2 mg/dL

(0.14 mmol/L)

5.0 mg/dL

(0.28 mmol/L)

1.0%

EDTA

5.0 mg/dL

(0.17 mmol/L)

-2.0 mg/dL

(-0.11 mmol/L)

-2.4%

Ephedrine

40 mg/dL

(2.42 mmol/L)

-3.9 mg/dL

(-0.22 mmol/L)

2.4%

Galactose

20 mg/dL

(1.11 mmol/L)

-3.1 mg/dL

(-0.17 mmol/L)

0.5%

Gentisic acid

7 mg/dL

(0.45 mmol/L)

7.2 mg/dL

(0.40 mmol/L)

2.9%

Glutathione

1 mg/dL

(0.03 mmol/L)

-2.6 mg/dL

(-0.14 mmol/L)

-3.7%

Haemoglobin

300 mg/dL

(0.05 mmol/L)

-3.1 mg/dL

(-0.17 mmol/L)

-2.6%

Heparin

2.1 mg/dL

(0.0018 mmol/L)

-3.0 mg/dL

(-0.17 mmol/L)

-1.3%

Ibuprofen

50 mg/dL

(2.43 mmol/L)

-2.6 mg/dL

(-0.15 mmol/L)

-1.9%

Icodextrin

1094 mg/dL (0.64~0.78 mmol/L)

-4.17 mg/dL

(-0.23 mmol/L)

-2.9%

Blood glucose 

value

Influence

Содержание GL 50

Страница 1: ...3 IN 1 BLOOD GLUCOSE MONITOR Step by step GL 50 EN...

Страница 2: ...ce for taking a sample 16 5 3 Taking a blood sample and measuring the glucose level 17 5 4 Reading the result and labelling measurements 19 5 5 Post processing and disposal 20 5 6 Evaluating measured...

Страница 3: ...lucose measurement of fresh whole blood samples either for self testing or in a clinical environment by trained personnel It enables you to measure your blood glucose quickly and easily store the meas...

Страница 4: ...ervice address A 3 in 1 monitor lancing device plug in USB and monitor B USB cover C 10 test strips D 10 sterile lancet needles 5 x 28G blue 5 x 33G purple E 2 button cell batteries 3 V CR2032 already...

Страница 5: ...ing at home The monitor enables you to quickly and simply Measure your blood glucose level Display labelled and save measured values Display the average measured blood glucose value for the last 7 14...

Страница 6: ...Maximum shelf life after initial opening in months Unit of measurement for blood glucose value LOT Batch designation Biohazard risk of infection Warning see accompanying documents Sterilised by radia...

Страница 7: ...Do not share the lancing device or lancet needles with others or among multiple patients risk of infection Use a new sterile lancet needle for each blood sample for single use only General notes Warn...

Страница 8: ...ended for measuring capillary and venous whole blood Storage and maintenance Warning Store the monitor and its accessories out of the reach of small children and pets Small parts such as lancet needle...

Страница 9: ...ecautions for handling blood when disposing of materials from the monitor Dispose of all blood samples and materials with which you or your patients come into contact correctly in order to prevent inj...

Страница 10: ...ce 2 Plug in USB 3 Monitor Front Rear 1 On Off button 2 Display 3 Rocker switch 4 Test strip slot 1 Battery cover 3 2 Lancing device and lancet needles 1 Cap 2 Protective lancet disc 3 Sterile lancet...

Страница 11: ...average blood glucose value ERR USB 7 Symbols for labelling measurements 8 Memory symbol 9 Test strip and blood droplet symbol 10 Blood glucose unit mg dL 11 Blood glucose unit mmol L Note To read the...

Страница 12: ...ne patient Handling test strips Note Securely close the test strip box immediately after taking out a test strip Do not use the test strips if they have expired The use of expired test strips may lead...

Страница 13: ...ettings 4 1 Removing the battery insulation strips replacing the batteries Note Two batteries are included in delivery of the blood glucose monitor These have already been inserted into the battery co...

Страница 14: ...ctly with a date and time and access them again later The time is displayed in the 24 hour format Set the year calendar to 2099 by pressing the or button Confirm by pressing the On Off button The day...

Страница 15: ...ch as the palm of the hand forearm or upper arm We recommend taking blood samples from the fingertip To make the procedure as painless as possible do not take samples directly from the centre of the f...

Страница 16: ...shed further into the holder 3 Remove the protective lancet disc Remove the protective lancet disc by turning it to a horizontal position Retain the protective disc for the safe disposal of the used l...

Страница 17: ...th tissue fluid and this may lead to an incorrect measurement Please note that insufficient blood circulation at the puncture area e g caused by cold temperatures or illness can lead to incorrect resu...

Страница 18: ...he fingertip Firmly position the lancing device slightly to the side of the centre of the fingertip Press the trigger Remove the lancing device from the finger A round drop of blood of at least 0 6 mi...

Страница 19: ...evice but no blood is added to the test strip within 2 minutes the device switches itself off Briefly remove the test strip and reinsert it so that the device automatically switches itself back on Con...

Страница 20: ...as it is displayed Once the value disappears from the display it can no longer be labelled 2 Press the rocker switch repeatedly Pressing once adds the label Before meals Pressing again adds the label...

Страница 21: ...k on the device 5 6 Evaluating measured blood glucose values Your blood glucose monitor can process measured values between 20 and 630 mg dL 1 1 and 35 0 mmol L The warning is displayed for measured v...

Страница 22: ...apillary plasma glucose 80 130 mg dL 4 4 7 2 mmol L Peak value of postprandial capillary plasma glucose 180 mg dL 10 0 mmol L For individual patients more or less strict glycaemic targets may be appro...

Страница 23: ...he test is being performed correctly Perform the control solution test if you suspect that the blood glucose monitor and or the test strips could be faulty or if you have repeatedly measured unusual b...

Страница 24: ...nt to save the control measurement in the memory press the rocker switch or again disappears from the display 4 Drip control solution onto surface Choose a clean surface to carry out the functional te...

Страница 25: ...cose monitor is too warm or too cold Bring the control solution test strip and monitor to room temperature 20 C to 26 C and repeat the test The check performed at room temperature is used as a general...

Страница 26: ...od glucose value for the last 7 14 30 and 90days Note If you have already saved measured values and you reset the date the average values are calculated from the new period indicates an empty memory f...

Страница 27: ...value and the average value are displayed this means 07 7 d days 2 Press repeatedly to display the average value for 7 14 30 and 90 days 3 You can cancel the process at any time To do so press the On...

Страница 28: ...t for blood glucose values and are shown on the display Press repeatedly to display the average blood glucose value for the last 14 30 and 90 days for values measured before meals After displaying the...

Страница 29: ...ucoMemory blood glucose evaluation software is pre installed on the monitor s USB stick You do not need to install the software locally on a PC This software enables you to evaluate your saved measure...

Страница 30: ...ions for use Remove the batteries if you do not intend to use the device for a prolonged period of time Only clean the device when it is switched off 7 1 Maintenance Clean the surface of the device us...

Страница 31: ...atteries empty Replace all batteries Ht Temperature of the measuring environment blood glucose monitor or test strip above the permitted range Repeat the test using a new test strip as soon as the mea...

Страница 32: ...Firmly insert the test strip contacts first into the slot on the device Make sure that the front of the test strip is facing you see Test strips on page 12 Device faulty Contact Customer Services Prob...

Страница 33: ...on Approx 5 seconds Calibration Plasma Testing method Amperometric bio sensor Usage Suitable for self testing System function test Each time device is switched on The serial number is located on the d...

Страница 34: ...on is similar to a blood test However control solution is used instead of blood The measurement using control solution must be within the result range This value range is printed on every test strip b...

Страница 35: ...4 3 2 1 0 1 2 7 2 134 0 7 4 2 5 0 1 1 9 3 338 1 18 8 8 1 0 4 2 4 System accuracy The GL50 blood glucose monitor in comparison to the YSI Three lots of GL50 blood glucose test strips have been tested t...

Страница 36: ...6 within 15 of the values obtained in the medical laboratory with glucose concentrations of at least 100 mg dL 5 55 mmol L You will find further details and information regarding blood glucose results...

Страница 37: ...amine 2 2 mg dL 0 14 mmol L 5 0 mg dL 0 28 mmol L 1 0 EDTA 5 0 mg dL 0 17 mmol L 2 0 mg dL 0 11 mmol L 2 4 Ephedrine 40 mg dL 2 42 mmol L 3 9 mg dL 0 22 mmol L 2 4 Galactose 20 mg dL 1 11 mmol L 3 1 m...

Страница 38: ...olazamide 2 5 mg dL 0 08 mmol L 2 3 mg dL 0 13 mmol L 1 8 Triglyceride 800 mg dL 9 37 mmol L 7 50 mg dL 0 42 mmol L 4 0 Uric acid 16 5 mg dL 0 98 mmol L 6 6 mg dL 0 37 mmol L 1 8 Xylose 9 5 mg dL 0 63...

Страница 39: ...claim shall only be considered if the buyer can provide Beurer or an authorised Beurer partner with a copy of the invoice purchase receipt and the original product The following are explicitly exclude...

Страница 40: ...ncetas lancette SteriLance Medical Suzhou Inc No 168 PuTuoShan Road New District Suzhou 215153 China GL 50 Beurer GmbH S flinger Stra e 218 89077 Ulm Germany www beurer com www beurer healthguide com...

Отзывы: