19
Classification, Symbols, Tables
TEST
LOCATION
Measurement
Uncertainty (dBuV)
Conducted Emissions 9kHz - 150kHz
Test Chamber A
3.61
Conducted Emissions 150kHz - 30MHz
Test Chamber A
3.23
Radiated Emissions 30MHz - 1GHz at
3 Meter/10 Meter
Open Area Test Site B
4.57
Conducted Emissions 9kHz - 150kHz
Open Area Test Site B
3.61
Conducted Emissions 150kHz - 30MHz
Open Area Test Site B
3.05
Chomerics Test Services - AVERAGE
Conducted Emissions 150kHz - 30MHz 9kHz Resolutions Bandwidth
FCC Part 15 Subpart B - Class B - Hot Line - AVERAGE
Frequency
Hot Line
Correction
AVG
Margin (dB)
Pass/Fail
Class B
MHz
AVG(dB)
Factor (dB)
dB(uV)
(+/-)
Indicator
Limit
0.47583
4.2
19.9
24.2
-22.5
PASS
47
0.47597
4.5
19.9
24.4
-22.2
PASS
47
0.48591
2.8
19.9
22.8
-23.6
PASS
46
0.49005
2.9
19.9
22.9
-23.4
PASS
46
0.49005
2.9
19.9
22.9
-23.4
PASS
46
0.4934
2.6
19.9
22.6
-23.6
PASS
46
Chomerics Test Services
Conducted Emissions 150kHz - 30MHz 9kHz Resolution Bandwidth
FCC Part 15 Subpart B - Class B - Hot Line - Quasi Peak
Frequency
Hot Line
Correction
QP
Margin (dB)
Pass/Fail
Class B
MHz
QP(dB)
Factor (dB)
dB(uV)
(+/-)
Indicator
Limit
0.47583
15.9
19.9
35.8
-20.9
PASS
57
0.47597
15.9
19.9
35.8
-20.9
PASS
57
0.48591
15.1
19.9
35
-21.4
PASS
56
0.49005
14.6
19.9
34.5
-21.8
PASS
56
0.49005
14.6
19.9
34.5
-21.8
PASS
56
0.49005
14.2
19.9
34.1
-22.1
PASS
56
Содержание SoClean 2