Chomerics Test Services
Conducted Emissions Class B 150kHz - 30 MHz 9kHz Resolution Bandwidth
FCC Part 15 Subpart B - Class B - Hot Line - AVERAGE
Frequency
Neutral
Line
Correction
AVG
Margin (dB)
Pass/Fail
Class B
MHz
AVG(dB)
Factor (dB)
dB(uV)
(+/-)
Indicator
Limit
0.47405
2.5
19.9
22.4
-24.4
PASS
47
0.48193
3
19.9
23
-23.5
PASS
47
0.48718
2.3
19.9
22.2
-24.1
PASS
46
0.48928
3
19.9
22.9
-23.4
PASS
46
0.49694
1.8
19.9
21.8
-24.3
PASS
46
0.50266
1.6
19.9
21.6
-24.4
PASS
46
Chomerics Test Services
Conducted Emissions Class B 150kHz - 30 MHz 9kHz Resolution Bandwidth
FCC Part 15 Subpart B - Class B - Neutral Line - Quasi Peak
Frequency
Hot Line
Correction
QP
Margin (dB)
Pass/Fail
Class B
MHz
QP(dB)
Factor (dB)
dB(uV)
(+/-)
Indicator
Limit
0.47405
14.3
19.9
34.2
-22.6
PASS
57
0.48193
14.2
19.9
34.1
-22.4
PASS
57
0.48718
13.9
19.9
33.8
-22.6
PASS
56
0.48928
14
19.9
33.9
-22.4
PASS
56
0.49694
13.5
19.9
33.4
-22.6
PASS
56
0.50266
13.1
19.9
33
-23