Important: user should make this setting before following operations. 

B.  In "Common Calibration"  interface, short press" 

.A.

11 

or 

"'f//f 

II 

to select gas, long press 

"OK"  button,  the  background of  selected  gas  will  turn  red,  connect  calibration  gas 

and after the value is stable,  long press "OK"  button to complete this operation with 
display of "Set OK". Short press "C) 

11 

to exit. 

Notice: 

a. Before calibration, activate this device for at least 6 minutes to allow it run stably
b. When the device is alarming for low power, stop calibration and charge it right away

c. Calibration gas must be standard gas used for calibration purpose 

d. Zero sensors firstly before calibration

e. Make sure gas flow is stable before connecting calibration gas, calibration gas flow for CH4,
CO, 02 is 200ml/min, 300ml/min for H2S.

f. Suggested calibration gases are as following:

CH4 1.00%vol, CO SOOppm, H2S S0ppm, oxygen can be calibrated in fresh air

11. In  setting  interface,  short  press  ,, .A."  or  ,, 

'f//f 

"  to  select  among "CH4", "LEL", "PPM",

"MG/m

3  ", 

"Allow  Calibration", "  ✓ 

11 

means  item  is selected. Long press "OK"  button to

select and confirm. Short press "<9 

11 

to exit.

Important: User can't enter calibration interface without choosing "Allow Calibration" 

4. Technical Parameter:

Technical 

Parameter 

Specification 

Measured gas 

Methane{CH4), Oxygen(O2), 

Carbon Monoxide{CO), Hydrogen Sulfide{H2S) 

Methane (CH4) 

Measurement range 

(o.oo~s.oo)%vol 

Measurement error 

(o.00~1.00)%vol:  ± 0.10%CH4 

(1.00~3.00)%vol: True value± 10% 

{3.00-5.00)%vol: 

±0.30%CH4 

Respond time 

�20s (T90) 

Resolution 

1.00%vol 

(Adjustable 

for 

full 

0.01%vol 

Alarm point 

range) 

Oxygen (02) 

Measurement range 

(o.o~3o.o)%vol 

Measurement error 

(o.o~s.o)%vol: 02: ±0.5%vol 

(> s.o~3o.o)%vol: O2:0.9%vol 

Respond time 

�35s (T90) 

Resolution 

0.1%vol 

18.0%vol{Adjustable for full range) 

Alarm point 

25.0%vol{Adjustable for full range) 

Carbon Monoxide (CO) 

Measurement range 

(0~1ooo)ppm 

Measurement error 

(0~2o)ppm: ± 2ppm 

(>20~10o)ppm:  ±4ppm 

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

 TestEquipmentDepot.com 

Содержание BST-MG08

Страница 1: ...t brand B Low Alarm L C High Alarm HJ D STEL Alarm SJ E TWA Alarm T F Preset Calibration Value G Product number and next calibrate date 30 seconds countdown H Auto Zero option BESANTEK LLJLIU Al i Rll SJ T ll ii Ml 0 00 02 20 9 l l OL 0 co Pl M H2S PPM 0 2Q16 09 22 II i II Alarm 9 Time Calibr ation Setting LOGO Main Interface Function interface 3 In auto zero option after seeing Adjust Zero long p...

Страница 2: ...h Alarm Set STEL Alarm and Set TWA Alarm Short press OK button to enter each setting Alarm L Alarm H STEL S TWA T short press A or T to set the value long press OK button to complete the operation with display set OK Short press 0 to exit 9 In time setting interface short press OK button to select Year Month Day Hour Min then press A or T to adjust the value Long press OK button to complete this o...

Страница 3: ...e calibrated in fresh air 11 In setting interface short press A or f f to select among CH4 LEL PPM MG m3 Allow Calibration 11 means item is selected Long press OK button to select and confirm Short press 9 11 to exit Important User can t enter calibration interface without choosing Allow Calibration 4 Technical Parameter Technical Parameter Specification Measured gas Methane CH4 Oxygen O2 Carbon M...

Страница 4: ...ifetime CH4 1 year CO H2S O2 2 years Alarms Audible buzzer with 85dB lm audible Visible flashing red LED 20m visible Vibration Charging time 4 Shours Continuous operating 13 hours hours Battery model PL123450 3 7V 1500mA Protect Grade IP66 Working temperature 2s c so c Size mm 130 70 37 Data logging for Continuously data logging MG08 only 100000 records with 1 10 minutes interval Enclosure PC ABS ...

Страница 5: ...cators user can know power left easily Technical parameter Samplin2 Air Flow 1600ml min Samplin2 Distance 30m Workin2 Temperature 20 500 Workin2 Humidity 90 RH Power Source 3 7V rechargeable lithium batte1y Output SW Input DCSV IA CharITTn2 Time 8h Weight 200g How to turn ON and OFF this pump Long press power switch to activate this pump and long press power switch again to shut it off ...

Страница 6: ...work 2 Gascheck Driver and Software Installation 2 1 Open CD Rom Driver on your computer and put the offered CD into it 2 2 Read the CD and find Gas Check Driver programme then choose light file according to your computer OS 2 3 Double click Prelnstaller exe or CP210xVCPinstaller exe and click Install then you ll see window prompt with Installation Successfully 2 4 To confilm this diiver has been ...

Страница 7: ...y device time needed is 1 6 minutes device will show Loading during this period 2 Check analyze these data on Data show section Histo1y data 3 This softwai e can get and store data from several gas detectors when data acquisition finished for one device click Disconnect and connect with another one 4 Click Clear data to delete all data collected by gas detector 5 When collecting data by device use...

Страница 8: ...values for each gas and each day Place mouse pointer on any point of collected data cmve you ll get exact time and concentration of this data Press left mouse and drag from left to light to spread all detailed da ta points and drag from 1ight to left to pull back or click cross point ofx and y axises to pull back On 1ight side of the cha1t you ll find 3 gas analyzing circle diagrams with Avg Min M...

Отзывы: