background image

-7-

Select the lowest test voltage (100V or 250V) range on the 

tester. Press the “test” button on the tester. Note the voltage 

LED (100V or 250V) on the TEST BOX. It should be lit with the 

same range on the tester. Record the reading on the tester. 

Repeat the same step on the other test voltage ranges.

NOTE:

 FOR THE LEDS TO BE PROPERLY LIT ON THE 

TEST BOX, THE RED TERMINAL MUST BE 

POSITIVE.ON SOME TESTERS THE RED LEAD IS 

NEGATIVE, THE LEADS MUST BE REVERSE.

B. CHECKING THE READING ACCURACY

Connect the tester to the 1M terminal on the TEST BOX as

shown above. Start with the lowest voltage range and record

the reading on the tester. Allow from 0.99 MΩ to 1.01 MΩ,

± 0.01MΩ.

Connect the tester to the 9.9M terminal on the TEST BOX.

Select the 500V test voltage range and record the reading on

the tester. Allow from 9.8 MΩ to 10.0 MΩ, ± 0.05MΩ.

Connect the tester to the 99M terminal on the TEST BOX.

Select the 1KV test voltage range and record the reading on

the tester.

Allow from 98 MΩ to 100 MΩ, ± 0.5MΩ.

Common Problems Solution

Only the lower Test

Voltage LED lit 

Low battery.

Tester reads over 

range all the time

Test leads open circuit. Note some 

testers require using the correct test 

leads.

Test Voltage LEDs 

do not light up

A: TEST BOX not plugged in.

B: Tester faulty or low battery.

C: Test leads open circuit, try to

short tester leads together and 

check zero.

D: Try to reverse test leads.

Содержание BST-CHK01

Страница 1: ...BOX Functional checking for RCD testers Loop testers and Insulation Continuity testers INSTRUCTION MANUAL 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Страница 2: ...Checking for the correct applied test voltage 6 7 Checking the reading accuracy 7 Checking the continuity measurement 8 9 Checking for the AC voltage measurement function 9 10 6 Checking a LOOP tester 11 12 7 Limited warranty 12 13 Due to our policy of constant improvement and development we reserve the right to change specifications without notice ...

Страница 3: ...t box while it is wet Keep dry at all times 5 Subjecting the test box to shock vibration and dropping may cause damage 6 Take all necessary safety precautions to ensure that there is no risk of testing before using 7 Safety standard EN 61010 1 EN 61326 1 IMPORTANT NOTE Dangerous voltages are present on the main sockets testers which have damaged cables etc Should not be tested before being repaire...

Страница 4: ...lation Testing Range 1MΩ 9 9MΩ 99MΩ Accuracy 1 Max Testing 1 2kV Continuity Testing Range 0 5Ω 2Ω 10Ω Accuracy 0 05Ω 0 1Ω 0 2 Ω Max Current 300mA LOOP Testing Range Supply Loop Supply Loop 1Ω Accuracy 0 45Ω 0 05Ω 1 45Ω 0 05Ω 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Страница 5: ...nt Temperature 0 C 40 C Observe the international electrical symbols listed below ri il Meter is protected throughout by double insulation or L J reinforced insulation Warning Risk of electric shock Caution Refer to this manual before using the meter N Alternating current IMPORTANT NOTE The earth line of TEST BOX is only a measuring line which is without any grounded protection 3 ...

Страница 6: ...e a 30mA RCD and the faulty range of your tester is on the 300mA range this will open any trip up to 300mA Test with both a long and a short trip time to ensure that it is correct It is less important if the tester has a positive negative cycle switch This will be better for timing measurement The unique features of the electronic trip of TEST BOX allow all the above to be tested quickly and easil...

Страница 7: ... lit but OVER CURRENT LED will not be lit Record the time displayed on the tester Allow from 135mS to 165mS 2mS Tripped is approximate to 3mAtolerance 10mA 3mA Over current is over 13mA Repeat the same step on the 30mA range of RCD tester The record of time shall be the same with the 10mA range Allow from 135mS to 165mS 2mS Tripped is approximate to 10mA tolerance 30mA 10mA Over current is over 40...

Страница 8: ...nprotected supply are bad connection A TEST BOX no plugged in RCD tester no power B Fuse blown on the right top of TEST BOX 5 Checking an installation or insulation tester The insulation tester uses a high voltage to measure a high resistance of Mega Ohm The TEST BOX must withstand the test without failure It is essential to measure both accuracy of the resistance and applied test voltage A CHECKI...

Страница 9: ...e and record the reading on the tester Allow from 0 99 MΩ to 1 01 MΩ 0 01MΩ Connect the tester to the 9 9M terminal on the TEST BOX Select the 500V test voltage range and record the reading on the tester Allow from 9 8 MΩ to 10 0 MΩ 0 05MΩ Connect the tester to the 99M terminal on the TEST BOX Select the 1KV test voltage range and record the reading on the tester Allow from 98 MΩ to 100 MΩ 0 5MΩ C...

Страница 10: ...eads shorted together and take this reading offall other readings AC IN 501 lz 210 250V Connect the test leads to the common and the 2 Ohm terminal as shown above Select the lowest range on the tester and record the reading on the tester Allow readings from 1 900 to 2 100 0 050 Connect the test leads to the common and the 10 Ohm terminal Select the lowest range on the tester and record the reading...

Страница 11: ...ll give a high zero Check tester is zeroed correctly with leads shorted together Tester reads over range The test leads are open circuit Note some test leads are fused and some use special connectors Check leads are OK D CHECKING FOR THE AC VOLTAGE MEASUREMENT FUNCTION Some multi function insulation and installation testers can measure AC voltage WARNING MAIN AC OUTPUT VOLTAGE IS PRESENT DURING TH...

Страница 12: ...below 1goo1 9 l t Jgoo l 0 0 t Q I g99 O 090 AC IN 50Hz 210 250V l23DVI 6 a Record voltage value displayed on meter Allow 210V to 250V Common Problems Solution A TEST BOX and unprotected supply have poor connection Check the No voltage displayed POWER LED on the center B Fuse blown on the right top of TEST BOX C Test leads ooen circuit 10 ...

Страница 13: ...PLUGGED INTO THE RCD SOCKET WHILE A LOOP TEST IS PERFORMING Loop test switch set on the SUPPLY LOOP position on TEST BOX Select the lowest range on the tester and press the test button on the loop tester Record the reading internal resistance approximate 0 45 0 05 Ohm on the tester Loop test switch set on the opposite position LOOP 1 R on TEST BOX Select the lowest range on the tester and press th...

Страница 14: ... are unable to repair or replace this product we will make a full refund of the purchase price Consult the user s manual for proper instruction regarding use of this instrument Our obligation under this warranty is limited to repairing replacing or making refund of this test equipment which proves to be defective within twelve months from the date of original purchase This warranty does not apply ...

Страница 15: ...e or misuse of the product by the purchaser or its user his employees or others and the remedies provided for in this warranty are expressly in lieu of any other liability we may have including incidental or consequential damages We reserve the right to discontinue models at any time or change specification price or design without notice and without incurring any obligation 99 Washington Street Me...

Отзывы: