Select the fitting range on the tester in order to read precise
measurement from each range on the tester.
For example, if the tester has a 200Ω, 20Ω and 2Ω ranges.
Use the 200Ω range to measure “10Ω”, the 20Ω to measure “2Ω”
and the 2Ω to measure “0.5Ω”.
Common Problems
Solution
Reading unstable
Check connections are clean and
tight in the sockets.
Reading high
Poor leads will give a high zero.
Check tester is zeroed correctly with
leads shorted together.
Tester reads over range
The test leads are open circuit. Note
some test leads are fused and some
use special connectors. Check leads
are OK.
D.
CHECKING FOR THE AC VOLTAGE MEASUREMENT
FUNCTION
Some multi-function, insulation and installation testers can
measure AC voltage.
WARNING: MAIN AC OUTPUT VOLTAGE IS PRESENT
DURING THIS TEST AND WATCH OUT FOR AVOIDING
ELECTRIC SHOCK.