Chapter 3 Parameters
46
Detector plateau measurement
Plateau measurement for detector function
(Only in the professional mode and only for NaI detector
and Super-Sens)
Start plateau measurement:
1
Push Enter to enable this function.
2
Push Cal. to start the plateau measurement.
The plateau measurement has been started. It may take
some minutes until the end of the plateau measurement
is reached.
First, the HV is set.
This may take several seconds.
The plateau measurement is carried out in
50V steps.
The count rate is averaged at each HV for 60
seconds each (can be set in code 37).
The HV starts at 500V and ends at the HV
set in code 67.
You have to push the Clear button twice to
abort a running plateau measurement.
After the plateau measurement is finished,
you can query the values by pushing the
buttons / .
The values remain stored until the next plat-
eau measurement is performed.
See also the section „Checking the Crystal-
Multiplier Combination“ in the hardware manual.
C o d e 6 8
Operating sequence of
a plateau measurement
500
63
550
179
600
205
1000
871
500
-
Содержание LB4710-050
Страница 2: ......
Страница 4: ......
Страница 8: ......
Страница 32: ......
Страница 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Страница 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Страница 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Страница 44: ......
Страница 58: ......
Страница 98: ......
Страница 104: ......
Страница 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Страница 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Страница 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Страница 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Страница 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Страница 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Страница 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Страница 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Страница 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Страница 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Страница 120: ......
Страница 141: ...Mini Switch LB 471 141 Notes...
Страница 142: ...Mini Switch LB 471 142 Notes...
Страница 145: ......
Страница 146: ...Limit Switch Mini Switch LB 471 Software Manual User s Guide detect and identify Process Control...
Страница 147: ......
Страница 149: ......
Страница 151: ......
Страница 161: ......
Страница 195: ......
Страница 215: ......
Страница 219: ......
Страница 223: ...Mini Switch LB 471 78 Notes...
Страница 224: ...Mini Switch LB 471 79 Notes...
Страница 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...