9.3 Changing alarm limits via SMS
Commands for setting alarm limits:
Setting
Status (on/off)
°C (00.0)
Duration in
minutes
aulimit=on,25.0,60
Upper alarm limit “on”
Temperature limit “25.0”
Time alarm “60”
allimit=on,15.0,60
Lower alarm limit “on”
Temperature limit “15.0”
Time alarm “60”
wulimit=off,0.0,0
Upper warning limit ”off”
Temperature limit “0.0” Time alarm “0”
wllimit=off,0.0,0
Lower warning limit “off”
Temperature limit “0.0” Time alarm “0”
Successful changes to the alarm limits are acknow-
ledged as
SUCCESSFUL
by
Fridge-tag
®
3
.
9.4 Adding alarm recipients
Admin sends a SMS with the command
sub41761234567
(mobile number of a new
recipient to the device). The device confirms with
SUCCESSFUL
.
A total of 5 different alarm recipients can be assigned.
Note
: A separate SMS must be sent for each alarm
recipient.
Conversation with +31 123456789 Thu., 09:26
sub41761234567
SUCCESSFUL
Fig. 9.4.1
Conversation with +31 123456789 Thu., 09:26
aulimit=on,25.0,60
SUCCESSFUL
Fig. 9.3.1
24