English
If not, the activation is now completed. Please con-
tinue with chapter 5 Placing the Fridge-tag 2 L.
Note:
During max. 1 minute after activation no temperature is
displayed on the screen.
If the device is confi gured with self-programmable
alarm limits proceed with the following chapter.
4.7) Setting the alarm limits
(Not standard, only by special order)
The fi rst digit is fl ashing:
Press READ to
adjust the number.
Press SET to
confi rm the number
1) and 3) Setting the HI & LO alarm duration, they
are completed in the same manner
This adjustment is done in 4 steps:
1) Setting the duration of the upper alarm limit
2) Setting the temperature of the upper alarm limit
3) Setting the duration to the lower alarm limit
4) Setting the temperature of the lower alarm limit
-27d
-25d
-23d
-21d
-19d
-17d
-15d
-13d
-1
1d
-9d
-7d
-5d
-3d
yesterd.
-29d
-
28d
-26d
-24d
-22d
-20d
-18d
-16d
-14d
-12d
-10d
-8d
-6d
-4d
-2d
today
Time / Alarm duration
SET
Alarm state
=OK
=ALARM
READ
®
Fridge-tag 2 L
Temperature (min/max)
<2°C
(15min)
>8°C
(30min)
225-0208-001-00E
110500000010
1
H
d r
u
0 00
1
1
2
The second digit is fl ashing:
Press READ to
adjust the number.
-27d
-25d
-23d
-21d
-19d
-17d
-15d
-13d
-1
1d
-9d
-7d
-5d
-3d
yesterd.
-29d
-
28d
-26d
-24d
-22d
-20d
-18d
-16d
-14d
-12d
-10d
-8d
-6d
-4d
-2d
today
Time / Alarm duration
SET
Alarm state
=OK
=ALARM
READ
®
Fridge-tag 2 L
Temperature (min/max)
<2°C
(15min)
>8°C
(30min)
225-0208-001-00E
110500000010
1
H
d r
u
0 00
2
1
2
Press SET to
confi rm the number
13