Operation and Menu Structure
NewSonic SonoDur-R
Issue 105 06/2020
Page 37
7.9.1.2
Adjustment by means of a new measured value recording
If the question "Use recorded readings for the adjustment?" was answered with "No", or if no
current measured values are available, the new measured values for adjustment must be included
in the submenu "Adjust measured value".
Thereafter, the target value must be changed to the reference value, as described above.
7.9.2
Adjusting number directly
If the adjustment number is already known, it can be entered directly by selecting "Direct adjustment
number". If a series of measurements is available, a question is raised again on whether a new series
of measurements should be started or not.
Figure 7.54
Figure 7.55
Figure 7.56
Figure 7.57
Figure 7.58
i
NOTE!
The procedure described in
Chapter 0
has the advantage that info key in the measurement menu
can be used to analyze the individual measured values in advance, so that any outliers can be
detected and eliminated. As a result, a "better" adjustment measurement is possible from the
outset, because this correction option for individual measurements is not provided otherwise.