Operation and Menu Structure
NewSonic SonoDur-R
Issue 105 06/2020
Page 58
Switch of the automatic output of the measuring values:
OFF:
<STX>#AO 0<CR><ETX>
Answer from Sono-R: <STX>#AO 0<CR><ETX>
ON:
<STX>#AO 1<CR><ETX>
Answer from Sono-R: <STX>#AO 1<CR><ETX>
Writing of the CAL-Faktors:
<STX>#CL 1234<CR><ETX>
Answer from Sono-R: <STX>#CL 1234<CR><ETX>
Answer from Sono-R, if the CAL-Faktor was given wrong: <STX>#CL ERR<CR><ETX>
Depending on the CAL-Factor, the current measuring row is being closed, saved automatically
under a standard-file name and the new measuring row will be stared.
Reading of the CAL-Faktors:
<STX>#CL ?<CR><ETX>
Answer from Sono-R: <STX>#CL 1234<CR><ETX>
Measuring start (only with Motor-Probe):
<STX>#MS<CR><ETX>
Answer from Sono-R: <STX>#MS<CR><ETX>
Answer from Sono-R, if measuring start is not possible (for Example whithin a current
measuring time): <STX>#MS ERR<CR><ETX>
Status request:
<STX>#ST<CR><ETX>
Answer from Sono-R, ready for the start of measuring : <STX>#ST RDY<CR><ETX>
Answer from Sono-R, current measurement: <STX>#ST MAC<CR><ETX>
Answer from Sono-R, measurment ended sucessfully, single output of the current taken
measurement value: <STX>#ST MEA 123.4 HRC<CR><ETX>
Answer from Sono-R, Motor drives up: <STX>#ST BSY<CR><ETX> (Only Motor-Probe)
Answer from Sono-R, Probe-Error (Probe disconnected, does not communicate): <STX>#ST
PROBE ERR<CR><ETX>
Answer from SonoDur-R, general Error: <STX>#ST ERR<CR><ETX>
Release of the File-Headers:
<STX>#FH<CR><ETX>
Answer from Sono-R: <STX>#FH<CR> File-Header in Text-Format as already given, lines with
<CR> separated <ETX>
Release of the last measuring value (repeatable):
<STX>#MW<CR><ETX>
Answer from SonoR-R: <STX>#MW 123.4 HRC<CR><ETX>
Answer from SonoR-R, if there is no measuring value given (only after the start of the
software, before the first measuring): <STX>#MW ERR<CR><ETX>
Release of the equipment-Softwareversion:
<STX>#GV<CR><ETX>
Answer from Sono-R: <STX>#GV 2.16<CR><ETX>